Literature DB >> 17111336

Endoscopic retrograde cholangiopancreatography catheter and accessory exchange using a short hydrophilic guide wire: a prospective study.

G I Papachristou1, T H Baron, F Gleeson, M J Levy, M D Topazian.   

Abstract

BACKGROUND AND STUDY AIMS: Although hydrophilic guide wires can be used to facilitate stricture cannulation during endoscopic retrograde cholangiopancreatography (ERCP), some endoscopists avoid using them because of concerns about wire loss during exchange. There are no data available on the outcomes of using a short hydrophilic guide wire during ERCP. The aim of this study was to assess the outcomes of therapeutic ERCP procedures in which a short, completely hydrophilic guide wire was used exclusively. PATIENTS AND METHODS: A total of 100 patients undergoing ERCP were studied prospectively. A 0.035-inch, 260-cm long, angled-tip hydrophilic wire (Terumo Glidewire) was used initially. Hydraulic catheter exchange was performed as follows: during catheter withdrawal the assistant advanced all the available wire into the catheter; a 12-ml syringe was then attached to the catheter and water was flushed under pressure to "float the wire" and maintain its position during catheter removal. Variables evaluated included exchange times and wire loss rates.
RESULTS: A total of 223 catheter exchanges were performed, 132 (59%) using the Olympus V-Scope (which held the wire in 62% of cases): 15% of exchanges were with catheters/accessories designed for short-wire use (Boston Scientific Rapid Exchange Biliary System or Rx System), and 85% were with a variety of standard-length accessories. Overall, the mean exchange time was 26 seconds (range 6 - 90 seconds, standard deviation 12 seconds). The mean exchange time was faster with the V-Scope and with non-Rx-System accessories. Wire loss occurred in 5 % of all exchanges. Desired ductal/stricture access was achieved in all the patients.
CONCLUSIONS: Exchange of short hydrophilic wires is quick and reliable. The Olympus V-Scope is able to hold the wire in some cases. Monorail-type devices and accessories slow catheter exchange down slightly because hydraulic exchange cannot be performed using these systems.

Entities:  

Mesh:

Year:  2006        PMID: 17111336     DOI: 10.1055/s-2006-944739

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  5 in total

1.  A new guidewire cannulation technique in ERCP: successful deep biliary access with triple-lumen sphincterotome and guidewire controlled by the endoscopist.

Authors:  Antonio López; Isabel Ferrer; Rosa Ana Villagrasa; Inmaculada Ortiz; Nuria Maroto; Cristina Montón; Joaquín Hinojosa; Eduardo Moreno-Osset
Journal:  Surg Endosc       Date:  2010-12-07       Impact factor: 4.584

Review 2.  ERCP wire systems: the long and the short of it.

Authors:  Shilpa Chandrupatla Reddy; Peter V Draganov
Journal:  World J Gastroenterol       Date:  2009-01-07       Impact factor: 5.742

3.  A prospective randomized study of thin versus regular-sized guide wire in wire-guided cannulation.

Authors:  Jorma Halttunen; Leena Kylänpää
Journal:  Surg Endosc       Date:  2012-12-13       Impact factor: 4.584

4.  Conventional endoscopic retrograde cholangiopancreaticography vs the Olympus V-scope system.

Authors:  Martin Raithel; Andreas Nägel; Jürgen Maiss; Dane Wildner; Alexander Fritzkarl Hagel; Sandra Braun; Hiwot Diebel; Eckhart Georg Hahn
Journal:  World J Gastroenterol       Date:  2013-03-28       Impact factor: 5.742

5.  EUS-guided biliary rendezvous using a short hydrophilic guidewire.

Authors:  Vinay Dhir; Boon Eu Andrew Kwek; Suryaprakash Bhandari; Mukta Bapat; Amit Maydeo
Journal:  J Interv Gastroenterol       Date:  2011-10-01
  5 in total