Experimental evaluation of a MOSFET dosimeter for proton dose measurements.

Ryosuke Kohno1, Teiji Nishio, Tomoko Miyagishi, Eriko Hirano, Kenji Hotta, Mitsuhiko Kawashima, Takashi Ogino.   

Abstract

The metal oxide semiconductor field-effect transistor (MOSFET) dosimeter has been widely studied for use as a dosimeter for patient dose verification. The major advantage of this detector is its size, which acts as a point dosimeter, and also its ease of use. The commercially available TN502RD MOSFET dosimeter manufactured by Thomson and Nielsen has never been used for proton dosimetry. Therefore we used the MOSFET dosimeter for the first time in proton dose measurements. In this study, the MOSFET dosimeter was irradiated with 190 MeV therapeutic proton beams. We experimentally evaluated dose reproducibility, linearity, fading effect, beam intensity dependence and angular dependence for the proton beam. Furthermore, the Bragg curve and spread-out Bragg peak were also measured and the linear-energy transfer (LET) dependence of the MOSFET response was investigated. Many characteristics of the MOSFET response for proton beams were the same as those for photon beams reported in previous papers. However, the angular MOSFET responses at 45, 90, 135, 225, 270 and 315 degrees for proton beams were over-responses of about 15%, and moreover the MOSFET response depended strongly on the LET of the proton beam. This study showed that the angular dependence and LET dependence of the MOSFET response must be considered very carefully for quantitative proton dose evaluations.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17110771     DOI: 10.1088/0031-9155/51/23/009

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  Dosimetric evaluation of a MOSFET detector for clinical application in photon therapy.

Authors:  Ryosuke Kohno; Eriko Hirano; Teiji Nishio; Tomoko Miyagishi; Tomonori Goka; Mitsuhiko Kawashima; Takashi Ogino
Journal:  Radiol Phys Technol       Date:  2007-11-17

2.  Investigation of an implantable dosimeter for single-point water equivalent path length verification in proton therapy.

Authors:  Hsiao-Ming Lu; Greg Mann; Ethan Cascio
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

3.  Evaluation of the usefulness of a MOSFET detector in an anthropomorphic phantom for 6-MV photon beam.

Authors:  Ryosuke Kohno; Eriko Hirano; Satoshi Kitou; Tomonori Goka; Kana Matsubara; Satoru Kameoka; Taeko Matsuura; Takaki Ariji; Teiji Nishio; Mitsuhiko Kawashima; Takashi Ogino
Journal:  Radiol Phys Technol       Date:  2010-02-04

4.  In vivo proton dosimetry using a MOSFET detector in an anthropomorphic phantom with tissue inhomogeneity.

Authors:  Ryosuke Kohno; Kenji Hotta; Kana Matsubara; Shie Nishioka; Taeko Matsuura; Mitsuhiko Kawashima
Journal:  J Appl Clin Med Phys       Date:  2012-03-08       Impact factor: 2.102

5.  Proton dose distribution measurements using a MOSFET detector with a simple dose-weighted correction method for LET effects.

Authors:  Ryosuke Kohno; Kenji Hotta; Taeko Matsuura; Kana Matsubara; Shie Nishioka; Teiji Nishio; Mitsuhiko Kawashima; Takashi Ogino
Journal:  J Appl Clin Med Phys       Date:  2011-04-04       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.