Biomechanical study on pullout strength of thoracic extrapedicular screw fixation.

Chang-feng Fu1, Yi Liu, Shao-kun Zhang, Zhi-ming Song.   

Abstract

OBJECTIVE: To identify the biomechanical feasibility of the thoracic extrapedicular approach to the placement of screws.
METHODS: Five fresh adult cadaveric thoracic spine from T1 to T8 were harvested. The screw was inserted either by pedicular approach or extrapedicular approach. The result was observed and the pullout strength by pedicular screw approach and extrapedicular screw approach via sagittal axis of the vertebrale was measured and compared statistically.
RESULTS: In thoracic pedicular approach, the pullout strength of pedicle screw was 1001.23 N+/-220 N (288.2-1561.7 N)ls and that of thoracic extrapedicular screw approach was 827.01 N+/-260 N when screw was inserted into the vertebrae through transverse process, and 954.25 N+/-254 N when screw was inserted into the vertebrae through the lateral cortex of the pedicle. Compared with pedicular group, the pullout strength in extrapedicular group was decreased by 4.7% inserted through transverse process (P larger than 0.05) and by 17.3% inserted through the lateral cortex (P less than 0.05). The mean pullout strength by extrapedicular approach was decreased by 11.04% as compared with pedicular approach (P less than 0.05).
CONCLUSIONS: It is feasible biomechanically to use extrapedicular screw technique to insert pedicular screws in the thoracic spine when it is hard to insert by pedicular approach.

Entities:  

Mesh:

Year:  2006        PMID: 17096934

Source DB:  PubMed          Journal:  Chin J Traumatol        ISSN: 1008-1275


  2 in total

1.  Comparison of pullout strength of the thoracic pedicle screw between intrapedicular and extrapedicular technique: a meta-analysis and literature review.

Authors:  Hua Wang; Huafeng Wang; Shilabant Sen Sribastav; Fubiao Ye; Chunxiang Liang; Zemin Li; Jianru Wang; Hui Liu; Xin Wang; Zhaomin Zheng
Journal:  Int J Clin Exp Med       Date:  2015-12-15

2.  Diagonal Trajectory Posterior Screw Instrumentation for Compromised Bone Quality Spine: Groove-Entry Technique/Hooking Screw Hybrid.

Authors:  Ivan Sekiguchi; Naoki Takeda; Naoki Ishida
Journal:  Spine Surg Relat Res       Date:  2018-04-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.