Literature DB >> 17095941

Tympanic temperature measurements: are they reliable in the critically ill? A clinical study of measures of agreement.

John L Moran1, John Victor Peter, Patricia J Solomon, Bernadette Grealy, Tania Smith, Wendy Ashforth, Megan Wake, Sandra L Peake, Aaron R Peisach.   

Abstract

OBJECTIVE: Accurate measurement of temperature is vital in the intensive care setting. A prospective trial was performed to compare the accuracy of tympanic, urinary, and axillary temperatures with that of pulmonary artery (PA) core temperature measurements.
DESIGN: A total of 110 patients were enrolled in a prospective observational cohort study.
SETTING: Multidisciplinary intensive care unit of a university teaching hospital. PATIENTS: The cohort was (mean +/- sd) 65 +/- 16 yrs of age, Acute Physiology and Chronic Health Evaluation (APACHE) II score was 25 +/- 9, 58% of the patients were men, and 76% were mechanically ventilated. The accuracy of tympanic (averaged over both ears), axillary (averaged over both sides), and urinary temperatures was referenced (as mean difference, Delta degrees centigrade) to PA temperatures as standard in 6,703 recordings. Lin concordance correlation (pc) and Bland-Altman 95% limits of agreement (degrees centigrade) described the relationship between paired measurements. Regression analysis (linear mixed model) assessed covariate confounding with respect to temperature modes and reliability formulated as an intraclass correlation coefficient.
MEASUREMENTS AND MAIN RESULTS: Concordance of PA temperatures with tympanic, urinary, and axillary was 0.77, 0.92, and 0.83, respectively. Compared with PA temperatures, Delta (limits of agreement) were 0.36 degrees C (-0.56 degrees C, 1.28 degrees C), -0.05 degrees C (-0.69 degrees C, 0.59 degrees C), and 0.30 degrees C (-0.42 degrees C, 1.01 degrees C) for tympanic, urinary, and axillary temperatures, respectively. Temperature measurement mode effect, estimated via regression analysis, was consistent with concordance and Delta (PA vs. urinary, p = .98). Patient age (p = .03), sedation score (p = .0001), and dialysis (p = .0001) had modest negative relations with temperature; quadratic relationships were identified with adrenaline and dobutamine. No interactions with particular temperature modes were identified (p > or = .12 for all comparisons) and no relationship was identified with either mean arterial pressure or APACHE II score (p > or = .64). The average temperature mode intraclass correlation coefficient for test-retest reliability was 0.72.
CONCLUSION: Agreement of tympanic with pulmonary temperature was inferior to that of urinary temperature, which, on overall assessment, seemed more likely to reflect PA core temperature.

Entities:  

Mesh:

Year:  2007        PMID: 17095941     DOI: 10.1097/01.CCM.0000250318.31453.CB

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  21 in total

1.  The mean prehospital machine; accurate prehospital non-invasive blood pressure measurement in the critically ill patient.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2010-06-08       Impact factor: 2.502

2.  The usefulness of an earphone-type infrared tympanic thermometer during cardiac surgery with cardiopulmonary bypass: clinical report.

Authors:  Taishi Masamune; Masanori Yamauchi; Keiichi Wada; Hironobu Iwashita; Katsumi Okuyama; Hirofumi Ino; Michiaki Yamakage; Tadahiko Ishiyama; Takashi Matsukawa
Journal:  J Anesth       Date:  2011-04-26       Impact factor: 2.078

3.  Fever is associated with delayed ventilator liberation in acute lung injury.

Authors:  Giora Netzer; David W Dowdy; Thelma Harrington; Satish Chandolu; Victor D Dinglas; Nirav G Shah; Elizabeth Colantuoni; Pedro A Mendez-Tellez; Carl Shanholtz; Jeffrey D Hasday; Dale M Needham
Journal:  Ann Am Thorac Soc       Date:  2013-12

Review 4.  Design, analysis, and interpretation of method-comparison studies.

Authors:  Sandra K Hanneman
Journal:  AACN Adv Crit Care       Date:  2008 Apr-Jun

5.  Design and optimization of an ultra wideband and compact microwave antenna for radiometric monitoring of brain temperature.

Authors:  Dario B Rodrigues; Paolo F Maccarini; Sara Salahi; Tiago R Oliveira; Pedro J S Pereira; Paulo Limao-Vieira; Brent W Snow; Doug Reudink; Paul R Stauffer
Journal:  IEEE Trans Biomed Eng       Date:  2014-04-15       Impact factor: 4.538

Review 6.  Hypothermia for acute brain injury--mechanisms and practical aspects.

Authors:  H Alex Choi; Neeraj Badjatia; Stephan A Mayer
Journal:  Nat Rev Neurol       Date:  2012-02-28       Impact factor: 42.937

7.  Body temperature mapping in critically ill newborn infants nursed under radiant warmers during intensive care.

Authors:  G K Chaseling; Y Molgat-Seon; T Daboval; S Chou; O Jay
Journal:  J Perinatol       Date:  2016-02-25       Impact factor: 2.521

8.  Ethnic differences in resting metabolic rate, respiratory quotient and body temperature: a comparison of Africans and European Australians.

Authors:  P A Adzika Nsatimba; K Pathak; M J Soares
Journal:  Eur J Nutr       Date:  2015-07-24       Impact factor: 5.614

9.  Non-invasive measurement of brain temperature with microwave radiometry: demonstration in a head phantom and clinical case.

Authors:  Paul R Stauffer; Brent W Snow; Dario B Rodrigues; Sara Salahi; Tiago R Oliveira; Doug Reudink; Paolo F Maccarini
Journal:  Neuroradiol J       Date:  2014-02-24

10.  Temperature rhythm in aneurysmal subarachnoid hemorrhage.

Authors:  Catherine J Kirkness; Robert L Burr; Hilaire J Thompson; Pamela H Mitchell
Journal:  Neurocrit Care       Date:  2008       Impact factor: 3.210