The size 1 LMA-ProSeal: Comparison with the LMA-Classic during pressure controlled ventilation in a neonatal intubation manikin.

M Micaglio1, D Trevisanuto, N Doglioni, G Zanette, V Zanardo, C Ori.   

Abstract

BACKGROUND: The classic laryngeal mask airway (cLMA) has been demonstrated to be effective for airway management during neonatal resuscitation. However, high airway pressures, when required, cannot be achieved with this device. A neonatal prototype of the LMA-ProSeal (PLMA), which might improve the oropharyngeal leak pressure, has recently been produced. The airway sealing pressures of the cLMA and the PLMA were compared in a neonatal manikin.
METHODS: A neonatal PLMA and a neonatal cLMA were positioned at random in a neonatal intubation manikin (Neonate Airway Trainer; Laerdal, Norway). A Dräeger pressure controlled ventilator (Dräeger 8000; Dräegerwerk AG, Germany) was connected to the airway tubes and increasing inspiratory pressures (from 10 to 40 cm H2O) of positive pressure ventilation applied. The peak and the mean airway pressures obtained with each device were recorded.
RESULTS: The airway pressures obtained with PLMA were significantly higher than those obtained with cLMA (p < 0.01) at levels of positive pressure ventilation of 25, 30, 35 and 40 cm H2O.
CONCLUSIONS: The neonatal PLMA allows higher airway pressure ventilation than the cLMA, in a neonatal intubation manikin. If confirmed clinically, this may have important implications during neonatal resuscitation when high airway pressures are required.

Entities:  

Mesh:

Year:  2006        PMID: 17095135     DOI: 10.1016/j.resuscitation.2006.06.026

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  3 in total

1.  Neonatal airway simulators, how good are they? A comparative study of physical and functional fidelity.

Authors:  T Sawyer; T P Strandjord; K Johnson; D Low
Journal:  J Perinatol       Date:  2015-11-19       Impact factor: 2.521

2.  Performance of size 1 I-gel compared with size 1 ProSeal laryngeal mask in anesthetized infants and neonates.

Authors:  Gulay Erdogan Kayhan; Zekine Begec; Mukadder Sanli; Ender Gedik; Mahmut Durmus
Journal:  ScientificWorldJournal       Date:  2015-02-22

3.  Supraglottic airway device versus a channeled or non-channeled blade-type videolaryngoscope for accidental extubation in the prone position: A randomized crossover manikin study.

Authors:  Hiroyuki Oshika; Yukihide Koyama; Masataka Taguri; Koichi Maruyama; Go Hirabayashi; Shoko Merrit Yamada; Masashi Kohno; Tomio Andoh
Journal:  Medicine (Baltimore)       Date:  2018-06       Impact factor: 1.889

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.