Inter-instrument agreement and influence of central corneal thickness on measurements with Goldmann, pneumotonometer and noncontact tonometer in glaucomatous eyes.

Viney Gupta1, Parul Sony, Harish C Agarwal, Ramanjit Sihota, Ajay Sharma.   

Abstract

PURPOSE: This study was conducted to compare the intraocular pressure (IOP) measurements by the Goldman applanation tonometer (GAT), non contact tonometer (NCT) and the ocular blood flow (OBF) pneumotonometer in different IOP ranges in glaucomatous eyes. The effect of central corneal thickness (CCT) on IOP measurement in chronic glaucomatous eyes using the three different tonometers was also evaluated.
MATERIALS AND METHODS: IOP measurements of 130 eyes of primary glaucoma patients were performed using GAT by an ophthalmologist while NCT and OBF-pneumotonometer measurements were performed by an experienced optometrist. The IOP values were compared amongst the three instruments in the three different IOP ranges (0-18 mmHg, > 18 to 25 mmHg, > 25 mmHg). CCT was also measured in all patients.
RESULTS: The mean of paired difference between GAT and NCT was 0.9 +/- 3.1 mmHg while that between GAT and OBF-pneumotonometer was 0.3 +/- 3.4 mmHg. The OBF-pneumotonometer and NCT were more affected by corneal thickness (0.41 mmHg and 0.4 mmHg / 10 micro corneal thickness respectively) while GAT was the least affected by corneal thickness (0.3 mmHg / 10 micro corneal thickness) though the difference was not statistically significant ( P =0.42).
CONCLUSION: With appropriate correction for corneal thickness the NCT and OBF-pneumotonometer can be used as reliably as GAT in following up glaucomatous patients.

Entities:  

Mesh:

Year:  2006        PMID: 17090879     DOI: 10.4103/0301-4738.27952

Source DB:  PubMed          Journal:  Indian J Ophthalmol        ISSN: 0301-4738            Impact factor:   1.848


  5 in total

1.  Clinical comparison of Pulsair non-contact tonometer and Goldmann applanation tonometer in Indian population.

Authors:  Shalini Mohan; Satyaprakash Tiwari; Arvind Jain; Jaya Gupta; Surendra Kumar Sachan
Journal:  J Optom       Date:  2013-07-26

2.  Performance of the PT100 noncontact tonometer in healthy eyes.

Authors:  Turki M AlMubrad
Journal:  Clin Ophthalmol       Date:  2011-05-19

3.  Detecting IOP Fluctuations in Glaucoma Patients.

Authors:  Brenda Nuyen; Kaweh Mansouri
Journal:  Open Ophthalmol J       Date:  2016-02-29

4.  The Association of Central Corneal Thickness and Intraocular Pressure Measures by Non-Contact Tonometry and Goldmann Applanation Tonometry among Glaucoma Patients.

Authors:  Samuel Kyei; Frank Assiamah; Michael Agyemang Kwarteng; Cynthia Pakyennu Gboglu
Journal:  Ethiop J Health Sci       Date:  2020-11

5.  Agreement between two Goldmann type applanation tonometers.

Authors:  Ronnie George; Hemamalini Arvind; M Baskaran; S Ve Ramesh; Prema Raju; Lingam Vijaya
Journal:  Indian J Ophthalmol       Date:  2008 Nov-Dec       Impact factor: 1.848

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.