Literature DB >> 17083439

Clinically relevant issues related to preheating composites.

Márcia Daronch1, Frederick A Rueggeberg, Linda Moss, Mario Fernando de Goes.   

Abstract

STATEMENT OF THE PROBLEM: Issues regarding the use of composite preheating need to be investigated so that the clinician will better understand the variables associated with this method.
PURPOSE: To examine the multiple aspects of use of a commercial composite preheating device (Calset, AdDent Inc., Danbury, CT, USA).
MATERIALS AND METHODS: Temperature values of three heating units and composite compules were obtained using a K-type thermocouple and were recorded digitally in real time. The following parameters were measured: maximum heater and composite temperature and its stability upon storage, composite temperature change when removed from the heater and injected, the effect of delivery system on ejected composite temperature, and the effect of repeated and extended preheating on composite monomer conversion (using infrared spectroscopy). Monomer conversion was measured after repeated composite cycling (from room temperature [RT] to 60 degrees C, 10x) or extended preheating (24 hours at 60 degrees C), and values were compared with composite maintained at RT (control group). Among test parameters, data (N=5 for each parameter) were analyzed using Student's t-test, analysis of variance, and the Tukey-Kramer post-hoc test where appropriate (alpha=0.05). RESULTS/
CONCLUSIONS: Two of the three tested units achieved the stated preset temperatures. Composite attained temperature values close to the heating unit. Composite temperature drop upon removal from the heater was dramatic: within 2 minutes a 50% temperature drop was noted. Heating the compule while preloaded in the syringe provided higher delivery temperatures than heating the compule separately (p < 0.00). Optimum results were achieved when preheated composite was dispensed and used as quickly as possible. Neither repeated nor extended preheating of composite significantly affected monomer conversion. CLINICAL SIGNIFICANCE: Preheating composite has potential benefits, but should be used with knowledge of its limitations. Reheating of unused composite does not affect its degree of conversion, thus decreasing material waste. Heating of the composite preloaded in the delivery syringe enhances the temperature of extruded composite.

Mesh:

Substances:

Year:  2006        PMID: 17083439     DOI: 10.1111/j.1708-8240.2006.00046.x

Source DB:  PubMed          Journal:  J Esthet Restor Dent        ISSN: 1496-4155            Impact factor:   2.843


  33 in total

1.  Effects of Preheated Composite on Micro leakage-An in-vitro Study.

Authors:  Joshua Ng Chor Yang; James David Raj; Herald Sherlin
Journal:  J Clin Diagn Res       Date:  2016-06-01

2.  Effect of different composite modulation protocols on the conversion and polymerization stress profile of bulk-filled resin restorations.

Authors:  M C G Erhardt; M Goulart; R C Jacques; J A Rodrigues; C S Pfeifer
Journal:  Dent Mater       Date:  2020-05-20       Impact factor: 5.304

3.  The effect of composite resin preheating on marginal adaptation of class II restorations.

Authors:  Farideh Darabi; Reza Tayefeh-Davalloo; Seyedeh-Maryam Tavangar; Fereshteh Naser-Alavi; Mozhdeh Boorboo-Shirazi
Journal:  J Clin Exp Dent       Date:  2020-07-01

4.  Effects of Rinsing Water Temperature and Preheated Composites on Microleakage of Composite Restorations with Two Bonding Agents.

Authors:  Mohammad Reza Malekipour; Mehrdad Barekatain; Farzaneh Shirani; Samaneh Alaei
Journal:  Front Dent       Date:  2022-01-04

Review 5.  [Effect of preheating on the properties of resin composite].

Authors:  Jing Xue; Bi-Na Yang
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2019-12-01

6.  Influence of preheating on mechanical and surface properties of nanofilled resin composites.

Authors:  Ali-Atef Elkaffass; Radwa-Ibrahim Eltoukhy; Salwa-Abd-Elraof Elnegoly; Salah-Hassab Mahmoud
Journal:  J Clin Exp Dent       Date:  2020-05-01

7.  Effect of pre-heating composite resin on gap formation at three different temperatures.

Authors:  Nilabh Choudhary; Sharad Kamat; Tm Mangala; Mohan Thomas
Journal:  J Conserv Dent       Date:  2011-04

8.  Effect of precuring warming on mechanical properties of restorative composites.

Authors:  Kareem Nada; Omar El-Mowafy
Journal:  Int J Dent       Date:  2011-10-27

9.  Polymerization behavior and thermal characteristics of two new composites at five temperatures: refrigeration to preheating.

Authors:  Tahereh Sadat Jafarzadeh-Kashi; Marzieh Mirzaii; Mohmmad Erfan; Akbar Fazel; Solmaz Eskandarion; Vahid Rakhshan
Journal:  J Adv Prosthodont       Date:  2011-12-28       Impact factor: 1.904

10.  Effect of cavity preparation techniques and different preheating procedures on microleakage of class V resin restorations.

Authors:  Emine Sirin Karaarslan; Aslihan Usumez; Bora Ozturk; Mehmet Ata Cebe
Journal:  Eur J Dent       Date:  2012-01