Literature DB >> 17080376

Experimental evaluation of an automated endoscope reprocessor with in situ generation of peracetic acid for disinfection of semicritical devices.

Syed A Sattar1, Richard J Kibbee, Jason A Tetro, Tony A Rook.   

Abstract

OBJECTIVE: To evaluate the effectiveness of a high-level disinfection solution generated inside an endoscope processing system for decontaminating external and internal surfaces of experimentally contaminated heat-sensitive medical devices.
METHODS: The American Society for Testing and Materials Simulated-Use Test protocol (E1837-02), which incorporates a soil load in each inoculum, was used to evaluate the efficacy of the system when processing 4 common types of endoscopes contaminated separately with 5 types of nosocomial pathogens: Pseudomonas aeruginosa (ATCC 15442), spores of Clostridium difficile (ATCC 9689), a glutaraldehyde-resistant strain of Mycobacterium chelonae, a vancomycin-resistant strain of Enterococcus faecalis, and a methicillin-resistant strain of Staphylococcus aureus. Rinse solution samples from channels and from surfaces of the processed endoscopes were tested for any microbicidal residues.
RESULTS: For all organisms tested, the baseline level of contamination of the endoscopes ranged from 5 log(10) to greater than 7 log(10) at each external surface site and internal channel. All tests showed reductions in viability of the test organisms to undetectable levels. All rinse solution samples from external and internal sites of the endoscopes proved to be free of any residual microbicidal activity.
CONCLUSIONS: The endoscope reprocessor, with its processor-generated high-level disinfection solution, successfully reduced the numbers of selected, clinically relevant pathogens to undetectable levels both in the channels and on the outside surfaces of the 4 representative endoscopes tested in this study.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17080376     DOI: 10.1086/508830

Source DB:  PubMed          Journal:  Infect Control Hosp Epidemiol        ISSN: 0899-823X            Impact factor:   3.254


  2 in total

Review 1.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

2.  Disinfective process of strongly acidic electrolyzed product of sodium chloride solution against Mycobacteria.

Authors:  Tomoyo Matsushita Yamamoto; Takashi Nakano; Masaki Yamaguchi; Mitsuhide Shimizu; Hong Wu; Hiroaki Aoki; Rie Ota; Toyohide Kobayashi; Kouichi Sano
Journal:  Med Mol Morphol       Date:  2012-12-07       Impact factor: 2.309

  2 in total