Cord and placenta arterial gas analysis: the accuracy of delayed sampling.

Adrienne Lynn1, Philip Beeby.   

Abstract

AIM: To determine the accuracy of delayed arterial gas sampling (1) from the umbilical cord and (2) from the placental surface at room temperature.
METHODS: Term deliveries were classified a priori into three groups: normal vaginal deliveries, elective caesarean sections and high risk deliveries. The cord was double clamped and paired arterial samples were taken from the cord and the placenta at 0, 30, 60 and 90 min.
RESULTS: 90 placentas were sampled with 30 cases per group. At time 0 the mean cord pH 7.207 (+/-0.08) was significantly lower than the placenta pH 7.240 (+/-0.08). The cord pH dropped significantly: by 0.050 (95% CI 0.036 to 0.063) at 30 min, 0.087 (95% CI 0.069 to 0.105) at 60 min, and 0.112 (95% CI 0.086 to 0.138) at 90 min. The placenta pH fell at twice the rate of the cord pH over 90 min. At time 0 the mean cord base excess -7.0 mmol/l (+/-4.1) was significantly lower than the placenta base excess -6.3 mmol/l (+/-3.6). The cord base excess fell at 30 min by 4.1 mmol/l (95% CI 3.4 to 4.7), at 60 min by 7.1 mmol/l (95% CI 6.1 to 8.0), and at 90 min by 9.0 mmol/l (95% CI 7.9 to 10.0). The pH and base excess rate of fall was similar for each of the three delivery groups despite differing starting values.
CONCLUSION: Arterial blood gases should be taken as soon as possible after delivery from the umbilical cord. However, when this is not possible, the arterial pH and base excess from a delayed sample from a clamped cord at room temperature can be used to estimate the values at birth.

Mesh:

Year:  2006        PMID: 17074787      PMCID: PMC2675428          DOI: 10.1136/adc.2006.103358

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  22 in total

Review 1.  A template for defining a causal relation between acute intrapartum events and cerebral palsy: international consensus statement.

Authors:  A MacLennan
Journal:  BMJ       Date:  1999-10-16

2.  Effect of delayed sampling on umbilical cord arterial and venous lactate and blood gases in clamped and unclamped vessels.

Authors:  L Armstrong; B Stenson
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2006-04-25       Impact factor: 5.747

Review 3.  Umbilical cord blood gas analysis at delivery.

Authors:  J A Thorp; G A Dildy; E R Yeomans; B A Meyer; V M Parisi
Journal:  Am J Obstet Gynecol       Date:  1996-09       Impact factor: 8.661

4.  Umbilical cord pH and base excess values in relation to adverse outcome events for infants delivering at term.

Authors:  Rahi Victory; Deborah Penava; Orlando Da Silva; Renato Natale; Bryan Richardson
Journal:  Am J Obstet Gynecol       Date:  2004-12       Impact factor: 8.661

5.  Do Apgar scores indicate asphyxia?

Authors:  G S Sykes; P M Molloy; P Johnson; W Gu; F Ashworth; G M Stirrat; A C Turnbull
Journal:  Lancet       Date:  1982-02-27       Impact factor: 79.321

6.  Umbilical blood gas analysis: I. Effect of storage of samples on outcome.

Authors:  V Q Nhan; H W de Bruyn; H J Huisjes
Journal:  Int J Gynaecol Obstet       Date:  1980 Mar-Apr       Impact factor: 3.561

7.  The reliability of the result of the umbilical cord pH.

Authors:  M Pel; P E Treffers
Journal:  J Perinat Med       Date:  1983       Impact factor: 1.901

8.  Placental compared with umbilical cord blood to assess fetal blood gas and acid-base status.

Authors:  Alison Nodwell; Lesley Carmichael; Michael Ross; Bryan Richardson
Journal:  Obstet Gynecol       Date:  2005-01       Impact factor: 7.661

9.  A fetal response to chorioamnionitis is associated with early survival after preterm birth.

Authors:  Monica M Lahra; Heather E Jeffery
Journal:  Am J Obstet Gynecol       Date:  2004-01       Impact factor: 8.661

10.  Sequential changes in placental blood gases and pH during the hour following delivery.

Authors:  J S Hilger; I R Holzman; D R Brown
Journal:  J Reprod Med       Date:  1981-06       Impact factor: 0.142

View more
  1 in total

Review 1.  Use of umbilical cord blood gas analysis in the assessment of the newborn.

Authors:  L Armstrong; B J Stenson
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2007-11       Impact factor: 5.747

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.