Literature DB >> 1706631

Elimination of electronic offset and physiological background activity in magnetocardiographic localization.

S Achenbach1, W Moshage, A Weikl, W Härer, K Abraham-Fuchs, K Göhl, K Bachmann.   

Abstract

A method has been developed to eliminate disturbing magnetic signals in the biomagnetic localization of arrhythmogenic sources in the heart. The procedure consists of two steps: Superimposed background activity of the heart is eliminated by subtraction of a template of pure background activity. Systematic and electronic offset is subsequently eliminated by baseline-correction during periods of zero activity. The method was applied to several kinds of arrhythmias. It was demonstrated that elimination of background activity is the prerequesite for exact localization and that the proposed procedure yields correct results.

Entities:  

Mesh:

Year:  1990        PMID: 1706631     DOI: 10.1515/bmte.1990.35.s3.160

Source DB:  PubMed          Journal:  Biomed Tech (Berl)        ISSN: 0013-5585            Impact factor:   1.411


  2 in total

1.  Clinical magnetocardiography: experience with a biomagnetic multichannel system.

Authors:  W Moshage; S Achenbach; A Weikl; K Göhl; K Bachmann; K Abraham-Fuchs; W Härer; S Schneider
Journal:  Int J Card Imaging       Date:  1991

2.  Evaluation of the non-invasive localization accuracy of cardiac arrhythmias attainable by multichannel magnetocardiography (MCG).

Authors:  W Moshage; S Achenbach; K Göhl; K Bachmann
Journal:  Int J Card Imaging       Date:  1996-03
  2 in total