Literature DB >> 17060162

Comparison of Goldmann applanation and applanation resonance tonometry in a biomicroscope-based in vitro porcine eye model.

P Hallberg1, K Santala, C Lindén, O A Lindahl, A Eklund.   

Abstract

We have developed an in vitro porcine eye model based on a biomicroscope, to simulate a clinical situation for IOP measurement on enucleated eyes. The aims of this study were to evaluate the model and to apply and compare Goldmann applanation tonometry (GAT) and applanation resonance tonometry (ART) measurements in porcine eyes. The GAT measurement (IOPGAT) showed a lower pressure, mean - 14.0 mm Hg (SD = 1.7 mm Hg) as compared with the reference pressure. For in vitro measurement with GAT on porcine eyes the linear calibration was IOP = 1.14 IOPGAT + 12.5 mm Hg (R2 = 0.99, p < 0.001, n = 280, four eyes). ART measurements correlated significantly to reference IOP, R = 0.86 (p < 0.001, n = 252, six eyes), with a mean difference of 5.4 mm Hg (SD = 6.7 mm Hg). GAT could only be used on porcine eyes if the IOP exceeded 13 mm Hg. Evaluation of the ART in this in vitro model showed position dependence for the sensor. To facilitate centre positioning a guiding tool is suggested. Porcine eyes are a possible substitute for human eyes in in vitro models for pre-clinical evaluation of new tonometry methods.

Entities:  

Mesh:

Year:  2006        PMID: 17060162     DOI: 10.1080/03091900512331332528

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  5 in total

1.  Underestimation of intraocular pressure after photorefractive keratectomy: a biomechanical analysis.

Authors:  Per Hallberg; Anders Eklund; Kenneth Santala; Timo Koskela; Olof Lindahl; Christina Lindén
Journal:  Med Biol Eng Comput       Date:  2006-07-26       Impact factor: 2.602

2.  Symmetric sensor for applanation resonance tomometry of the eye.

Authors:  Per Hallberg; Christina Lindén; Tomas Bäcklund; Anders Eklund
Journal:  Med Biol Eng Comput       Date:  2006-03       Impact factor: 2.602

3.  Can we trust intraocular pressure measurements in eyes with intracameral air?

Authors:  Gauti Jóhannesson; Christina Lindén; Anders Eklund; Anders Behndig; Per Hallberg
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-07-17       Impact factor: 3.117

4.  Repeatability and reproducibility of applanation resonance tonometry: a cross-sectional study.

Authors:  Laura Ottobelli; Paolo Fogagnolo; Paolo Frezzotti; Stefano De Cillà; Elena Vallenzasca; Maurizio Digiuni; Ruggiero Paderni; Ilaria Motolese; Simone Alex Bagaglia; Eduardo Motolese; Luca Rossetti
Journal:  BMC Ophthalmol       Date:  2015-04-10       Impact factor: 2.209

5.  Investigation of intraocular pressure of the anterior chamber and vitreous cavity of porcine eyes via a novel method.

Authors:  Koji Nagae; Hiromasa Sawamura; Makoto Aihara
Journal:  Sci Rep       Date:  2020-11-25       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.