Literature DB >> 17051096

Style 410 cohesive silicone breast implants: safety and effectiveness at 5 to 9 years after implantation.

Per Hedén1, Beate Boné, Diane K Murphy, Araceli Slicton, Patricia S Walker.   

Abstract

BACKGROUND: Inamed introduced a fourth-generation implant, the Style 410 cohesive silicone breast implant, in 1993. To date, little has been published on this implant despite its widespread use in Europe, and no data are available on its long-term rupture rate.
METHODS: Women implanted with Style 410 implants between 1995 and 1999 at Akademikliniken Hospital in Stockholm were eligible to participate in the study. One hundred forty-four of 302 potential subjects received a physical examination, underwent magnetic resonance imaging screening for silent rupture, and completed a quality-of-life survey. The majority of subjects were Caucasian (96 percent), with a median age of 39 years. The study population consisted of subjects with cosmetic augmentation (86 percent) or revision of previous breast implantation (14 percent). Median time elapsed between physical examination and magnetic resonance imaging was 2 days.
RESULTS: The 144 subjects provided 286 implants for magnetic resonance imaging assessment, with a median implantation time of 6 years (range, 5 to 9 years). Overall, 99.0 percent of implants showed no evidence of rupture, 0.3 percent showed evidence of rupture, and 0.7 percent were indeterminate. The most common complication noted at physical examination was capsular contracture (5.6 percent). All other complications occurred in less than 3 percent of subjects. Quality-of-life results found an improvement in overall sense of well-being for 87 percent of subjects. Most compelling is that 97 percent stated an overall feeling that their breast implantation had been advantageous.
CONCLUSIONS: The authors' findings support the long-term safety and effectiveness of Inamed Style 410 cohesive breast implants and demonstrate a low prevalence of rupture.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17051096     DOI: 10.1097/01.prs.0000239457.17721.5d

Source DB:  PubMed          Journal:  Plast Reconstr Surg        ISSN: 0032-1052            Impact factor:   4.730


  12 in total

1.  Textured Breast Implants: A Closer Look at the Surface Debris Under the Microscope.

Authors:  Leland H Webb; Victoria L Aime; Annie Do; Kenneth Mossman; Raman C Mahabir
Journal:  Plast Surg (Oakv)       Date:  2017-06-28       Impact factor: 0.947

2.  Influence of patient age on capsular contracture after aesthetic breast augmentation.

Authors:  Philip H Zeplin; Niamph Corduff
Journal:  Plast Surg (Oakv)       Date:  2015       Impact factor: 0.947

3.  A Case of Silicone and Sarcoid Granulomas in a Patient with "Highly Cohesive" Silicone Breast Implants: A Histopathologic and Laser Raman Microprobe Analysis.

Authors:  Todor I Todorov; Erik de Bakker; Diane Smith; Lisette C Langenberg; Linda A Murakata; Mark H H Kramer; Jose A Centeno; Prabath W B Nanayakkara
Journal:  Int J Environ Res Public Health       Date:  2021-04-24       Impact factor: 3.390

Review 4.  Capsular Contracture after Breast Augmentation: An Update for Clinical Practice.

Authors:  Hannah Headon; Adbul Kasem; Kefah Mokbel
Journal:  Arch Plast Surg       Date:  2015-09-15

5.  Surface Areas of Textured Breast Implants: Implications for the Biofilm Theory of Capsule Formation.

Authors:  Tim Brown
Journal:  Plast Reconstr Surg Glob Open       Date:  2018-03-19

6.  Toll-Like Receptors (TLRs) Expression in Contracted Capsules Compared to Uncontracted Capsules.

Authors:  Y Bachour; M J P F Ritt; R Heijmans; F B Niessen; S P Verweij
Journal:  Aesthetic Plast Surg       Date:  2019-04-01       Impact factor: 2.326

7.  Review of 1,447 breast augmentation patients using PERTHESE silicone implants.

Authors:  Jorge Miguel Psillakis; Paulo Henrique Facchina; Paulo Kharmandayan; Luis Trillo; Waldecir Chiarelo Canzi; Herberti Rosique Aguiar
Journal:  Aesthetic Plast Surg       Date:  2009-09-17       Impact factor: 2.326

8.  Silicone migration to the contralateral axillary lymph nodes and breast after highly cohesive silicone gel implant failure: a case report.

Authors:  Gabriel J Kaufman; Rita A Sakr; Cyrille Inguenault; Isabelle Sarfati; Claude Nos; Krishna B Clough
Journal:  Cases J       Date:  2009-03-10

9.  Long-term safety and effectiveness of style 410 highly cohesive silicone breast implants.

Authors:  Per Hedén; Giorgio Bronz; Jens Jorgen Elberg; Rika Deraemaecker; Diane K Murphy; Araceli Slicton; R James Brenner; Catarina Svarvar; Joost van Tetering; Leo Peter van der Weij
Journal:  Aesthetic Plast Surg       Date:  2009-05-13       Impact factor: 2.326

10.  One stage rescue procedure after capsular contracture of breast implants with autologous fat grafts collected by water assisted liposuction ("BEAULI Method").

Authors:  Klaus Ueberreiter; Ursula Tanzella; Felix Cromme; Dietrich Doll; Björn Dirk Krapohl
Journal:  GMS Interdiscip Plast Reconstr Surg DGPW       Date:  2013-01-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.