| Literature DB >> 17050326 |
J-C Lu1, H-R Xu, F Chen, Y-F Huang, N-Q Lu.
Abstract
We attempted to provide an instruction for the standardization of alpha-glucosidase level determination and quality controls within the same laboratory and among different laboratories. Each of 51 semen samples was divided into two aliquots, centrifuging at 1000 g for 10 min or 3000 g for 15 min. The alpha-glucosidase level and sperm concentration of each sample were assayed. The alpha-glucosidase level in seminal plasma obtained at 3000 g for 15 min centrifugation was significantly lower than that at 1000 g for 10 min (p = 0.001). An additional 6 samples of seminal plasma with or without phenylmethylsulfonyl fluoride (PMSF), obtained at 3000 g for 15 min centrifugation, were frozen for 20 days, and each of 6 samples was determined for their alpha-glucosidase levels after thawing every other day. There was no significant difference between alpha-glucosidase levels in seminal plasma regardless of the presence of PMSF. The alpha-glucosidase level increased with the length of abstinence period. In conclusion, centrifugal velocity and abstinence time should be standardized in the alpha-glucosidase determination. Frozen seminal plasma may serve as the sample for quality control among clinical laboratories.Entities:
Mesh:
Substances:
Year: 2006 PMID: 17050326 DOI: 10.1080/01485010600822705
Source DB: PubMed Journal: Arch Androl ISSN: 0148-5016