BACKGROUND: Our study is aimed to determine the performance of 3 automated urinalysis systems-Clinitek Atlas, Urisys 2400 and Aution Max.
METHODS: One thousand urine specimens were analyzed with the 3 automated systems. The results of the 3 assays were compared for testing urine chemistry and evaluating the capacity of leukocyte esterase and nitrite to detect bacteriuria.
RESULTS: The correlation between the 3 instruments represented as within 1 grading difference was better between the Atlas and Aution Max systems for pH, blood, glucose, urobilinogen, ketone and specific gravity. For protein and nitrite, better correlation was observed between the Atlas and Urisys 2400, while the Aution Max and Urisys 2400 conveyed better correlation for bilirubin and white blood cells. The sensitivity and specificity of both the leukocyte esterase and nitrite in screening for significant bacteriuria were 71.7, 58.9, 70.8% and 99.1, 99.1 and 97.2%, for the Clinitek Atlas, Aution Max and Urisys 2400, respectively.
CONCLUSIONS: The automated urinalysis systems demonstrate acceptable correlations with each other in urine chemistries, especially between the Clinitek Atlas and Aution Max systems on the majority of items. The specificity and negative predictive value of leukocyte esterase and nitrite of the 3 instruments for screening of significant bacteriuria were sufficient to avoid unnecessary urine culture.

Entities:  

Mesh:

Year:  2006        PMID: 17049339     DOI: 10.1016/j.cca.2006.08.033

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  6 in total

1.  Performance Evaluation of Three URiSCAN Devices for Routine Urinalysis.

Authors:  Kiwoong Ko; Min-Jung Kwon; Seungho Ryu; Hee-Yeon Woo; Hyosoon Park
Journal:  J Clin Lab Anal       Date:  2015-08-24       Impact factor: 2.352

2.  Risk of poor outcomes with novel and traditional biomarkers at clinical AKI diagnosis.

Authors:  Isaac E Hall; Steven G Coca; Mark A Perazella; Umo U Eko; Randy L Luciano; Patricia R Peter; Won K Han; Chirag R Parikh
Journal:  Clin J Am Soc Nephrol       Date:  2011-10-27       Impact factor: 8.237

3.  Analytical performance, agreement and user-friendliness of six point-of-care testing urine analysers for urinary tract infection in general practice.

Authors:  Marjolein J C Schot; Sanne van Delft; Antoinette M J Kooijman-Buiting; Niek J de Wit; Rogier M Hopstaken
Journal:  BMJ Open       Date:  2015-05-18       Impact factor: 2.692

4.  Case report: proximal tubule impairment following volatile anesthetic exposure.

Authors:  Evan C Ray; Khaled Abdel-Kader; Nicholas Bircher; Helbert Rondon-Berrios
Journal:  Physiol Rep       Date:  2015-09

5.  Prospective, observational study comparing automated and visual point-of-care urinalysis in general practice.

Authors:  Sanne van Delft; Annelijn Goedhart; Mark Spigt; Bart van Pinxteren; Niek de Wit; Rogier Hopstaken
Journal:  BMJ Open       Date:  2016-08-08       Impact factor: 2.692

6.  Diaper-embedded urine test device for the screening of urinary tract infections in children: a cohort study.

Authors:  Niko Paalanne; Lotta Wikstedt; Tytti Pokka; Jarmo Salo; Matti Uhari; Marjo Renko; Terhi Tapiainen
Journal:  BMC Pediatr       Date:  2020-08-11       Impact factor: 2.125

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.