Literature DB >> 17038139

Effect of percutaneous interventions within the coronary sinus on the success rate of the implantations of resynchronization pacemakers.

Oskar Kowalski1, Radosław Lenarczyk, Janusz Prokopczuk, Patrycja Pruszkowska-Skrzep, Teresa Zielińska, Beata Sredniawa, Agata Musialik-Łydka, Sławomir Pluta, Tomasz Kukulski, Mariola Szulik, Lech Poloński, Zbigniew Kalarus.   

Abstract

BACKGROUND: Cardiac resynchronization therapy (CRT) becomes a "gold standard" in therapy of selected patients with advanced heart failure. We set out to evaluate the feasibility and safety of percutaneous interventions within coronary sinus (CS) and their effect on the success rate of left ventricular (LV) lead implantation during CRT.
METHODS: The study analyzed eight consecutive patients with the indications for CRT, who needed additional procedures within CS to overcome technical problems during left ventricular (LV) electrode implantation. The analyzed group consisted of three subgroups: patients in whom percutaneous balloon angioplasty within CS was needed (n = 4); patients with acute instability of the lead, requiring stenting of the vein to fix the electrode (n = 2); and patients with the stenting of CS due to late dislocation of the lead (n = 2). Success rate, procedure duration, fluoroscopy, complications, and electrical parameters of leads were analyzed.
RESULTS: Success rate of the procedures was 87.5%; additional interventions increased overall efficacy of CRT implantation at our center from 88% to 98% (P < 0.05). Procedure duration (155.0 minute) and fluoroscopy time (42.5 minute) remained acceptable for the patient and operator; however, both were higher than in the procedures performed routinely in our hospital. Electrical properties of the LV leads were stable and within normal ranges during the observation period. We noted two local dissections of CS during the procedure, which remained clinically silent.
CONCLUSION: Percutaneous interventions within CS seem to be feasible and safe treatment options, which can improve the short- and long-term success rates of CRT.

Entities:  

Mesh:

Year:  2006        PMID: 17038139     DOI: 10.1111/j.1540-8159.2006.00501.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Different venous angioplasty manoeuvres for successful implantation of CRT devices.

Authors:  Guido Luedorff; Rainer Grove; Wolfgang Kranig; J Thale
Journal:  Clin Res Cardiol       Date:  2008-12-18       Impact factor: 5.460

2.  Utility of a novel pacing guidewire in pre-implantation testing at different left ventricular sites in cardiac resynchronization therapy procedures.

Authors:  Ngai Yin Chan; Chi Chung Choy; Chun Leung Lau; Ying Keung Lo; Pui Shan Chu; Ho Chuen Yuen; Suet Ting Lau
Journal:  J Interv Card Electrophysiol       Date:  2011-08-09       Impact factor: 1.900

3.  A streamlined technique of trans-septal endocardial left ventricular lead placement.

Authors:  Ernest W Lau
Journal:  J Interv Card Electrophysiol       Date:  2009-04-22       Impact factor: 1.900