R J Carreira1, F M Cordeiro, A J Moro, M G Rivas, R Rial-Otero, E M Gaspar, I Moura, J L Capelo.
Abstract
New findings in sample treatment based on high-intensity focused ultrasound (HIFU) for protein digestion after polyacrylamide gel electrophoresis separation are presented. The following variables were studied: (i) sample volume; (ii) sonotrode diameter; (iii) previous protein denaturation; (iv) cooling; (v) enzyme concentration; and (vi) protein concentration. Results showed that positive protein identification could be done after protein separation by gel electrophoresis through peptide mass fingerprint (PMF) in a volume as low as 25 microL. The time needed was less than 2 min and no cooling was necessary. The importance of the sonotrode diameter was negligible. On the other hand, protein denaturation before sonication was a trade-off for the success of procedure here described. The protein coverage was raised from 5 to 30%, and the number of peptides matching the proteins was also increased in a percentage ranging 10-100% when the classical overnight treatment is compared with the proposed HIFU procedure. The minimum amount of protein that can be identified using the HIFU sample treatment by matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF-MS) was 0.06 microg. The lower concentration of trypsin successfully used to obtain an adequate protein digestion was 3.6 microg/mL.Entities:
Mesh:
Substances:
Year: 2006 PMID: 17034802 DOI: 10.1016/j.chroma.2006.09.078
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759