| Literature DB >> 17027306 |
Thorsten Maly1, Jeff Bryant, David Ruben, Robert G Griffin.
Abstract
We describe a field-lock/field-sweep system for the use in superconducting magnets. The system is based on a commercially available field mapping unit and a custom designed broad-band 1H NMR probe. The NMR signal of a small water sample is used in a feedback loop to set and control the magnetic field to high accuracy. The current instrumental configuration allows field sweeps of +/-0.4 T and a resolution of up to 10(-5) T (0.1 G) and the performance of the system is demonstrated in a high-field electron paramagnetic resonance (EPR) application. The system should also be of utility in other experiments requiring precise and reproducible sweeps of the magnetic field such as DNP, ENDOR or PELDOR.Entities:
Mesh:
Year: 2006 PMID: 17027306 PMCID: PMC2518535 DOI: 10.1016/j.jmr.2006.09.012
Source DB: PubMed Journal: J Magn Reson ISSN: 1090-7807 Impact factor: 2.229