Wear and corrosion in retrieved thoracolumbar posterior internal fixation.

Marta L Villarraga1, Peter A Cripton, Stephanie D Teti, Duane L Steffey, Saki Krisnamuthy, Todd Albert, Alan Hilibrand, Alexander Vaccaro.   

Abstract

STUDY
DESIGN: Posterior thoracolumbar spine implants retrieved as part of routine clinical practice over a 2-year period were analyzed to identify wear and corrosion.
OBJECTIVE: Engineering analyses of retrieved posterior instrumentation for indications of performance and failure and correlation of this information with clinical factors. SUMMARY OF BACKGROUND DATA: Recent studies have reported spinal instrumentation particulate wear debris and have noted the importance of design considerations at implant connector interfaces.
METHODS: A total of 57 implants were analyzed from patients (39 female, 18 male) whose average age at implantation was 43.9 years (range, 13.7-77.4 years). Time of implantation ranged from 2 months to 13.5 years. The top 3 implantation diagnoses were radiculopathy (33%), scoliosis (30%), and back pain (25%). Metallurgical analyses were performed to characterize the wear and/or corrosion, and fractures of the implants.
RESULTS: Wear was present in 75%, corrosion in 39%, and fractures in 7% of the retrieved implants. Wear and/or corrosion was more prevalent, with respect to the total number of implants retrieved, in implants that had been in service at least 1 year. There was no evidence of corrosion in any of the Ti implants, whereas corrosion was present (with wear) in 58% of the stainless steel (SS) implants. Wear and corrosion were more frequently observed in long rods than in short rods. Implantation times were longer for SS implants than for Ti implants.
CONCLUSIONS: Retrieved rods exhibited corrosion, wear, and fracture, with wear and corrosion mainly located at the interfaces with hooks, screws, or cross-connectors. The mechanisms causing this material loss in situ, as well as what local or systemic responses it may stimulate are of clinical significance and should be studied further.

Entities:  

Mesh:

Year:  2006        PMID: 17023855     DOI: 10.1097/01.brs.0000239132.16484.be

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  6 in total

1.  Serum titanium, niobium and aluminium levels two years following instrumented spinal fusion in children: does implant surface area predict serum metal ion levels?

Authors:  Thomas P Cundy; William J Cundy; Georgia Antoniou; Leanne M Sutherland; Brian J C Freeman; Peter J Cundy
Journal:  Eur Spine J       Date:  2014-11       Impact factor: 3.134

2.  Corrosion of Harrington rod in idiopathic scoliosis: long-term effects.

Authors:  Beth Sherman; Tanya Crowell
Journal:  Eur Spine J       Date:  2017-06-17       Impact factor: 3.134

3.  Obstructed spinopelvic fixation in the setting of a triangular titanium sacroiliac fusion implant: a case description.

Authors:  Joseph L Laratta; James D Lin; Jamal N Shillingford; Nathan E Hardy; Hemant Reddy; Ronald A Lehman
Journal:  J Spine Surg       Date:  2017-12

4.  Prospective study on serum metal levels in patients with metal-on-metal lumbar disc arthroplasty.

Authors:  Matthew F Gornet; J K Burkus; M L Harper; F W Chan; A K Skipor; J J Jacobs
Journal:  Eur Spine J       Date:  2012-11-20       Impact factor: 3.134

5.  Chromium and Nickel Concentrations in Subjects with a Stainless Steel Metal-on-Metal Cervical Disc Arthroplasty: Results from a Prospective Longitudinal Study with 7 Years Follow-Up.

Authors:  Vaneet Singh; Anastasia K Skipor; Abdulhafez A Selim; Joshua J Jacobs
Journal:  Int J Spine Surg       Date:  2018-08-31

6.  Comparison of in vivo and simulator-retrieved metal-on-metal cervical disc replacements.

Authors:  Steven M Kurtz; Lauren Ciccarelli; Megan L Harper; Ryan Siskey; Jacob Shorez; Frank W Chan
Journal:  Int J Spine Surg       Date:  2012-12-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.