Literature DB >> 17011847

Optimization of anterior eye fluorescein viewing.

Rachael C Peterson1, James S Wolffsohn, Colin W Fowler.   

Abstract

PURPOSE: To optimize anterior eye fluorescein viewing and image capture.
DESIGN: Prospective experimental investigation.
METHODS: The spectral radiance of ten different models of slit-lamp blue luminance and the spectral transmission of three barrier filters were measured. Optimal clinical instillation of fluorescein was evaluated by a comparison of four different instillation methods of fluorescein into 10 subjects. Two methods used a floret, and two used minims of different concentration. The resulting fluorescence was evaluated for quenching effects and efficiency over time.
RESULTS: Spectral radiance of the blue illumination typically had an average peak at 460 nm. Comparison between three slit-lamps of the same model showed a similar spectral radiance distribution. Of the slit-lamps examined, 8.3% to 50.6% of the illumination output was optimized for >80% fluorescein excitation, and 1.2% to 23.5% of the illumination overlapped with that emitted by the fluorophore. The barrier filters had an average cut-off at 510 to 520 nm. Quenching was observed for all methods of fluorescein instillation. The moistened floret and the 1% minim reached a useful level of fluorescence in on average approximately 20s ( approximately 2.5x faster than the saturated floret and 2% minim) and this lasted for approximately 160 seconds.
CONCLUSIONS: Most slit-lamps' blue light and yellow barrier filters are not optimal for fluorescein viewing and capture. Instillation of fluorescein using a moistened floret or 1% minim seems most clinically appropriate as lower quantities and concentrations of fluorescein improve the efficiency of clinical examination.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17011847     DOI: 10.1016/j.ajo.2006.04.062

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  6 in total

1.  Prevalence of dry eye in video display terminal users: a cross-sectional Caucasian study in Italy.

Authors:  Gemma Caterina Maria Rossi; Luigia Scudeller; Federica Bettio; Gian Maria Pasinetti; Paolo Emilio Bianchi
Journal:  Int Ophthalmol       Date:  2018-06-07       Impact factor: 2.031

2.  Comparative Evaluation of Clinical Methods of Tear Film Stability Assessment: A Randomized Crossover Trial.

Authors:  Michael T M Wang; Jennifer P Craig
Journal:  JAMA Ophthalmol       Date:  2018-03-01       Impact factor: 7.389

3.  Analysis of the first tear film break-up point in Sjögren's syndrome and non-Sjögren's syndrome dry eye patients.

Authors:  Songjiao Zhao; Qihua Le
Journal:  BMC Ophthalmol       Date:  2022-01-03       Impact factor: 2.209

4.  Comparative of meibomian gland morphology in patients with evaporative dry eye disease versus non-dry eye disease.

Authors:  Ricaurte Ramiro Crespo-Treviño; Anna Karen Salinas-Sánchez; Francisco Amparo; Manuel Garza-Leon
Journal:  Sci Rep       Date:  2021-10-20       Impact factor: 4.379

5.  Impact of Bariatric Surgery-Induced Weight Loss on Anterior Eye Health in Patients with Obesity.

Authors:  Said Karimzad; Paramdeep S Bilkhu; James S Wolffsohn; Srikanth Bellary; Hala Shokr; Rishi Singhal; Doina Gherghel
Journal:  Nutrients       Date:  2022-06-14       Impact factor: 6.706

6.  Feasibility of Silicon Quantum Dots as a Biomarker for the Bioimaging of Tear Film.

Authors:  Sidra Sarwat; Fiona Jane Stapleton; Mark Duncan Perry Willcox; Peter B O'Mara; Richard David Tilley; J Justin Gooding; Maitreyee Roy
Journal:  Nanomaterials (Basel)       Date:  2022-06-08       Impact factor: 5.719

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.