Dosimetric verification for intensity-modulated radiotherapy of thoracic cancers using experimental and Monte Carlo approaches.

Si Young Jang1, H Helen Liu, Xiaochun Wang, Oleg N Vassiliev, Jeffrey V Siebers, Lei Dong, Radhe Mohan.   

Abstract

PURPOSE: To investigate the dosimetric accuracy of commercial treatment planning systems used in intensity-modulated radiotherapy (IMRT) for thoracic cancer. METHODS AND MATERIALS: Clinical IMRT plans for lung and esophageal cancers and mesothelioma were used to investigate the accuracy of dose calculations from two commercial treatment planning systems (Pinnacle and Corvus systems). Dose distributions were measured with ion chambers and thermoluminescent dosimeters for individual IMRT fields and composite treatment plans in water phantoms and anthropomorphic phantoms. A Monte Carlo-based system was established to compute three-dimensional dose distributions to compare with the treatment planning system calculations.
RESULTS: Dose calculations from the Pinnacle system were acceptable within 5% of the local dose or a 5-mm distance-to-agreement for 80% of the points measured with ion chambers, 74% of the points measured with thermoluminescent dosimeters, and 96% of the points compared with the Monte Carlo calculations. For the Corvus system, 89% of the points agreed with the measured dose and 98% agreed with the Monte Carlo calculations. Underestimation of the dose from the treatment planning system was found in the low-dose regions (<50% of the prescribed dose), possibly caused by inadequate modeling of the multileaf collimators.
CONCLUSION: The Pinnacle and Corvus dose calculations were acceptable for thoracic IMRT in high-dose regions. Beam modeling is likely the most critical factor for the accuracy of IMRT dose calculations.

Entities:  

Mesh:

Year:  2006        PMID: 17011466     DOI: 10.1016/j.ijrobp.2006.06.048

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  3 in total

1.  Current therapies for malignant pleural mesothelioma.

Authors:  Takashi Nakano
Journal:  Environ Health Prev Med       Date:  2008-02-28       Impact factor: 3.674

2.  Comparison of 2D and 3D gamma analyses.

Authors:  Kiley B Pulliam; Jessie Y Huang; Rebecca M Howell; David Followill; Ryan Bosca; Jennifer O'Daniel; Stephen F Kry
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

3.  Clinical relevance of different dose calculation strategies for mediastinal IMRT in Hodgkin's disease.

Authors:  J Koeck; Y Abo-Madyan; H T Eich; F Stieler; J Fleckenstein; J Kriz; R-P Mueller; F Wenz; F Lohr
Journal:  Strahlenther Onkol       Date:  2012-06-29       Impact factor: 3.621

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.