Evaluation of RF heating on humerus implant in phantoms during 1.5T MR imaging and comparisons with electromagnetic simulation.

Hiroyuki Muranaka1, Takayoshi Horiguchi, Shuji Usui, Yoshitake Ueda, Osamu Nakamura, Fumiaki Ikeda, Ken Iwakura, Giichirou Nakaya.   

Abstract

PURPOSE: To evaluate the effect of radiofrequency (RF) heating on a metallic implant during magnetic resonance imaging (MRI), temperatures at several positions of an implant were measured, and results are compared with electromagnetic simulations using a finite element method.
METHODS: A humerus nail implant made of stainless steel was embedded at various depths of tissue-equivalent gel-phantoms with loop (loop phantom) and partially cut loop (loop-cut phantom), and the phantoms were placed parallel to the static magnetic field of a 1.5T MRI device. Scans were conducted at maximum RF for 15 min, and temperatures were recorded with 2 RF-transparent fiberoptic sensors. Finally, electromagnetic-field analysis was performed.
RESULTS: Temperatures increased at both ends of the implants at various depths, and temperature increase was suppressed with increasing depth. The maximum temperature rise was 12.3 degrees C at the tip of the implant and decreased for the loop-cut phantom. These tendencies resembled the results of electromagnetic simulations.
CONCLUSION: RF heating was verified even in a nonmagnetizing metal implant in a case of excessive RF irradiation. Particularly, rapid temperature rise was observed at both ends of the implant having large curvatures. The difference in temperature increase by depth was found to reflect the skin-depth effect of RF intensity. Electromagnetic simulation was extremely useful for visualizing the eddy currents within the loop and loop-cut phantoms and for evaluating RF heating of a metallic implant for MRI safety.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17008764     DOI: 10.2463/mrms.5.79

Source DB:  PubMed          Journal:  Magn Reson Med Sci        ISSN: 1347-3182            Impact factor:   2.471


  3 in total

1.  Assessment of magnetic field interactions and radiofrequency-radiation-induced heating of metallic spinal implants in 7 T field.

Authors:  Itsuko Tsukimura; Hideki Murakami; Makoto Sasaki; Hirooki Endo; Daisuke Yamabe; Ryosuke Oikawa; Minoru Doita
Journal:  J Orthop Res       Date:  2017-03-08       Impact factor: 3.494

2.  Visible Human Project® female surface based computational phantom (Nelly) for radio-frequency safety evaluation in MRI coils.

Authors:  Gregory M Noetscher; Peter Serano; William A Wartman; Kyoko Fujimoto; Sergey N Makarov
Journal:  PLoS One       Date:  2021-12-10       Impact factor: 3.240

3.  Signal voids of active cardiac implants at 3.0 T CMR.

Authors:  Theresa Reiter; Ingo Weiss; Oliver M Weber; Wolfgang R Bauer
Journal:  Sci Rep       Date:  2022-04-15       Impact factor: 4.996

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.