Illumination correction on MR images.

Edoardo Ardizzone1, Roberto Pirrone, Orazio Gambino.   

Abstract

OBJECTIVE: An important artifact corrupting Magnetic Resonance Images is the rf inhomogeneity, also called bias artifact. This anomaly produces an abnormal illumination fluctuation on the image, due to variations of the device magnetic field. This artifact is particularly strong on images acquired with a device specialized on upper and lower limbs due to their coil configuration. A method based on homomorphic filtering aimed to suppress this artifact was proposed by Guillemaud. This filter has two faults: it doesn't provide an indication about the cutoff frequency (cf) and introduces another illumination artifact on the edges of the foreground. This work is an improvement to this method because it resolves both problems.
METHODS: The experimental setup has been performed on knee images obtained by 5 volunteers and acquired through an Artoscan device using the following parameters: Spin Echo sequence, Repetition time: 980 ms, Echo time: 26 ms, Slice thickness: 4 mm, Flip Angle: 90 degrees .
RESULTS: Two specialists in orthoptics evaluated the results of the proposed approach by examining the restored images and validating the results produced by the filter. A quantitative evaluation has been performed on a manually segmented restored image using the coefficient of variation (cv) measure.
CONCLUSIONS: Following the specialists qualitative evaluation, the illuminance of upper and lower peripheral zones results to be enhanced; a loose of contrast can be noted only in few cases. The Bias image exhibits an artifact focused usually on the central part of the foreground. The quantitative evaluation based on cv shows that this index is lowered for all the segmented regions with respect to the original value. The method is automatic and doesn't require any hypothesis on the tissues. A manual version of the algorithm can be also implemented allowing the physician to choose the preferred cf. In this case the value selected by the method can be considered as a default value.

Entities:  

Mesh:

Year:  2006        PMID: 17006728     DOI: 10.1007/s10877-006-9040-1

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  12 in total

1.  Adaptive fuzzy segmentation of magnetic resonance images.

Authors:  D L Pham; J L Prince
Journal:  IEEE Trans Med Imaging       Date:  1999-09       Impact factor: 10.048

2.  Qualitative and quantitative evaluation of six algorithms for correcting intensity nonuniformity effects.

Authors:  J B Arnold; J S Liow; K A Schaper; J J Stern; J G Sled; D W Shattuck; A J Worth; M S Cohen; R M Leahy; J C Mazziotta; D A Rottenberg
Journal:  Neuroimage       Date:  2001-05       Impact factor: 6.556

3.  Automated model-based bias field correction of MR images of the brain.

Authors:  K Van Leemput; F Maes; D Vandermeulen; P Suetens
Journal:  IEEE Trans Med Imaging       Date:  1999-10       Impact factor: 10.048

4.  Retrospective correction of MR intensity inhomogeneity by information minimization.

Authors:  B Likar; M A Viergever; F Pernus
Journal:  IEEE Trans Med Imaging       Date:  2001-12       Impact factor: 10.048

5.  A modified fuzzy C-means algorithm for bias field estimation and segmentation of MRI data.

Authors:  Mohamed N Ahmed; Sameh M Yamany; Nevin Mohamed; Aly A Farag; Thomas Moriarty
Journal:  IEEE Trans Med Imaging       Date:  2002-03       Impact factor: 10.048

6.  Polynomial modeling and reduction of RF body coil spatial inhomogeneity in MRI.

Authors:  M Tincher; C R Meyer; R Gupta; D M Williams
Journal:  IEEE Trans Med Imaging       Date:  1993       Impact factor: 10.048

7.  Correction of intensity variations in MR images for computer-aided tissue classification.

Authors:  B M Dawant; A P Zijdenbos; R A Margolin
Journal:  IEEE Trans Med Imaging       Date:  1993       Impact factor: 10.048

8.  Adaptive segmentation of MRI data.

Authors:  W M Wells; W L Grimson; R Kikinis; F A Jolesz
Journal:  IEEE Trans Med Imaging       Date:  1996       Impact factor: 10.048

9.  Optimized homomorphic unsharp masking for MR grayscale inhomogeneity correction.

Authors:  B H Brinkmann; A Manduca; R A Robb
Journal:  IEEE Trans Med Imaging       Date:  1998-04       Impact factor: 10.048

10.  Intensity correction in surface-coil MR imaging.

Authors:  L Axel; J Costantini; J Listerud
Journal:  AJR Am J Roentgenol       Date:  1987-02       Impact factor: 3.959

View more
  1 in total

1.  Image quality improvement of composed whole-spine MR images by applying a modified homomorphic filter--first results in cases of multiple myeloma.

Authors:  M P Lichy; C Mueller-Horvat; V Jellus; W Horger; M Horger; C Pfannenberg; B Kiefer; C D Claussen; H P Schlemmer
Journal:  Eur Radiol       Date:  2008-05-29       Impact factor: 5.315

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.