Abstract

AIM: In the presence of coronary artery disease, implantable cardioverter-defibrillators (ICD) are used effectively for treating life-threatening tachyarrhythmias. Continuous monitoring of myocardial ischaemia would provide a new diagnostic option in future ICD generations. METHODS AND
RESULTS: In 22 selected patients undergoing coronary angioplasty, percutaneous transluminal coronary angioplasty (PTCA), three electrodes, similar to those used in the ICD, were inserted aiming to create six intra-thoracic ECG (IT-ECG) leads according to Einthoven and Goldberger. In total, 27 PTCA were conducted. The diagnostic efficacy for ischaemia assessment was compared with the surface ECG. The IT-ECG proved to be more sensitive than conventional ECG in early and overall ischaemia assessment. At 30 s of coronary artery occlusion, ischaemic ST-segment alterations (> or =0.25 mV) were present in the IT-ECG 2.3 times more often (23 vs. 10/27 PTCA attempts, P<0.01) and at 90 s 1.4 times more often compared with conventional ECG leads (18 vs. 26/27, P<0.05). Intra-thoracic Einthoven 2 (SVC+RVA vs. ICD-housing) and Goldberger 3 (SVC+ICD-housing vs. RVA) had the highest sensitivity (88/85%). Using > or =4 IT-ECG, ischaemia monitoring was independent of severity and site of origin. IT-ECG signals showed double ST-T signal amplitude (4.19+/-0.6 vs. 2.15+/-0.3 mV, ratio: 1.95, P<0.01) at a QRS/ST amplitude ratio similar in the two ECG techniques.
CONCLUSION: This study provides strong evidence that the ICD-based IT 6-lead ECG would provide a new and efficient means of assessing a patient's daily ischaemic burden.

Entities:  

Mesh:

Year:  2006        PMID: 17005592     DOI: 10.1093/europace/eul104

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  1 in total

1.  Continuous ST-Monitoring Function of Implantable Cardioverter Defibrillator Detects Silent Ischemia in Patients With Coronary Artery Disease.

Authors:  Tetsuya Watanabe; Keiji Hirooka; Yoshio Furukawa; Masanori Yabuki; Akio Hirata; Kazunori Kashiwase; Ryu Shutta; Takanao Mine; Hiroya Mizuno; Toshikazu Tanaka; Takahiro Doi; Akihiro Yoshida; Yuji Okuyama; Shinsuke Nanto
Journal:  J Am Heart Assoc       Date:  2018-06-30       Impact factor: 5.501

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.