| Literature DB >> 16996759 |
Andrew E Marble1, Igor V Mastikhin, Bruce G Colpitts, Bruce J Balcom.
Abstract
The design and construction of a unilateral NMR (UMR) magnet assembly for near-surface 1D profiling is presented. The arrangement consists of a single permanent magnet topped with a shaped iron pole cap. The analytically determined profile of the pole cap shapes the field over the magnet, giving a constant gradient of 31 G/cm over a 8mm depth at a 1H frequency of 4.26 MHz in a spot approximately 5 mm wide. The moderate gradient allows 1D profiling of planar samples with a frequency encoded spin-echo experiment. The curvature of the magnetic field limits the available resolution to 100's of microm. The device is suitable for profiling planar samples in which a coarse resolution but large spatial extent is desired.Entities:
Mesh:
Year: 2006 PMID: 16996759 DOI: 10.1016/j.jmr.2006.08.013
Source DB: PubMed Journal: J Magn Reson ISSN: 1090-7807 Impact factor: 2.229