| Literature DB >> 16996325 |
Abstract
A bioanalytical method for determination of eflornithine (DFMO) in 1000 microL human plasma has been developed and validated. DFMO and the internal standard (IS) were analysed by liquid chromatography with evaporative light-scattering detection (ELSD). Separation was performed on a Chirobiotic TAG (250 mm x 4.6 mm) column with ethanol (99.5%):0.01 mol/L acetic acid-triethylamine buffer at the rate of 25:75% (v/v) with flow rate of 1.0 mL/min. For d-DFMO in plasma the inter-assay precision was 6.5% at 75 micromol/L, 6.6% at 375 micromol/L and 5.8% at 750 micromol/L. For l-DFMO in plasma the inter-assay precision was 10.4% at 75 micromol/L, 6.5% at 375 micromol/L and 5.0% at 750 micromol/L. The lower limit of quantification (LLOQ) was determined to 25 micromol/L where the precision was 4.3% and 5.7%, respectively.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16996325 DOI: 10.1016/j.jchromb.2006.08.030
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205