| Literature DB >> 16991204 |
Jorge J Soto-Chinchilla1, Ana M García-Campaña, Laura Gámiz-Gracia, Carmen Cruces-Blanco.
Abstract
A CZE method with UV-Vis detection has been established and validated for the determination of nine sulfonamides: sulfapyridine, sulfamethazine, sulfamerazine, sulfamether, sulfadiazine, sulfadimethoxine, sulfamethoxazole, sulfachlorpyridazine, and sulfamethizole. Optimum separation was obtained on a 64.5 cm x 75 microm bubble cell capillary using a buffer containing 45 mM sodium phosphate and 10% methanol at pH 7.3, with temperature and voltage of 27 degrees C and 25 kV, respectively. p-Aminobenzoic acid was used as an internal standard . Taking into account the lack of sensitivity of the UV-Vis detection, the application of an on-line preconcentration methodology, such as large-volume sample stacking with polarity switching has been proposed. This procedure combined with a solvent extraction/SPE method applied for off-line preconcentration and cleanup provides a significant improvement in the LODs, ranging from 2.59 to 22.95 mug/L for the studied compounds; the quantification of these residues being possible below the levels established by EU legislation in animal food products, such as meat. Satisfactory recoveries were also obtained in the analysis of these compounds in ground water.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16991204 DOI: 10.1002/elps.200600166
Source DB: PubMed Journal: Electrophoresis ISSN: 0173-0835 Impact factor: 3.535