Ventricular cross stimulation in a dual chamber pacing system: phenomenon analysis.

A Puglisi1, R Ricci, P Azzolini, C Peraldo Neja, M Fioranelli, G Speciale, G Angrisani.   

Abstract

Seven cases of ventricular cross stimulation from a group of 23 patients implanted with DDD devices are presented. In two patients the phenomenon was observed at the moment of DDD programming at nominal values, and in five other patients it was reproduced by increasing the atrial output voltage up to ten volts. In all 23 patients cross stimulation disappeared permanently within 24 hours after implantation. From the onset of cross stimulation to its end, atrial and ventricular threshold voltages were unchanged, while the atrial and ventricular impedances significantly decreased. These results suggest that an important role in the phenomenon occurs by impedance variation at the interface between the pulse generator and body tissue.

Entities:  

Mesh:

Year:  1990        PMID: 1697965     DOI: 10.1111/j.1540-8159.1990.tb02146.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Case of Dual Chamber Pacemaker with Cross Stimulation.

Authors:  R Bhardwaj; R Marwah
Journal:  Med J Armed Forces India       Date:  2011-07-21

2.  Syncope triggered by atrial flutter in a patient with a pacemaker: cross-stimulation-a case report.

Authors:  Olga Durán-Bobin; Francisco Crespo-Mancebo; Juliana Elices-Teja; Carlos González-Juanatey
Journal:  Eur Heart J Case Rep       Date:  2021-02-18