Literature DB >> 16972059

Long-term results of 58 hip cup revision arthroplasties using a threaded ring implant.

Christoph von Schulze Pellengahr1, Tobias Düll, Peter E Müller, Hans Roland Dürr, Andrea Baur-Melnyk, Markus Maier, Christof Birkenmaier, Volkmar Jansson.   

Abstract

INTRODUCTION: There is still a discussion whether cementless hip cup revisions should be performed with a press fit cup or a threaded ring implant.
MATERIALS AND METHODS: The results of 58 hip cup changes using the spherical, threaded ring "Munich" are presented. In 16 cases, the ring "Munich I" with a smooth surface and in 42 cases the ring "Munich II" with a corundum-blasted surface were implanted. Mean follow-up was 7.5 years; radiological signs of loosening and the Harris Hip Score were examined.
RESULTS: The re-revision rate for aseptic loosening of the ring was 12/58 overall (Munich I: 5/16, Munich II: 7/42). The average Harris Hip Score was 76.2 overall (Munich I: 74.4, Munich II: 76.8). Higher survival rates were achieved with the "Munich II" ring due to improved construction details.
CONCLUSION: These rates are still below those reported for press-fit cups, the technique described by Slooff or reinforcement rings.

Entities:  

Mesh:

Year:  2006        PMID: 16972059     DOI: 10.1007/s00402-006-0187-5

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  1 in total

1.  Long-term results of the Wagner conical screw cup.

Authors:  Alexander Schuh; Annemarie Schraml; Gerd Hohenberger
Journal:  Int Orthop       Date:  2008-06-26       Impact factor: 3.075