Literature DB >> 16970333

Passive sampling of airborne peroxyacetic acid.

Hartmut Henneken1, Laura Assink, Joyce de Wit, Martin Vogel, Uwe Karst.   

Abstract

The first passive sampling device for the determination of airborne peroxyacetic acid (PAA) is presented. 2-([3-{2-[4-Amino-2-(methylsulfanyl)phenyl]-1-diazenyl}phenyl]sulfonyl)-1-ethanol (ADS) is used to impregnate glass fiber filters, and the reagent is oxidized by PAA to the corresponding sulfoxide ADSO. After elution of the filters, ADS and ADSO are separated by reversed-phase HPLC and detected by UV/visible absorbance. Limit of detection is 30 ppb, limit of quantification is 90 ppb (for 30 min sampling), and the linear range comprises 2 orders of magnitude. Thorough investigations were carried out with respect to the selectivity of the method toward hydrogen peroxide, and air samples were analyzed successfully after disinfection of a laboratory area.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16970333     DOI: 10.1021/ac060668h

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  1 in total

1.  A field-portable colorimetric method for the measurement of peracetic acid vapors: a comparison of glass and plastic impingers.

Authors:  Angela L Stastny; Amos Doepke; Robert P Streicher
Journal:  J Occup Environ Hyg       Date:  2022-06-28       Impact factor: 3.359