Comparison of fluorescein and phycoerythrin conjugates for quantifying CD20 expression on normal and leukemic B-cells.

Lili Wang1, Fatima Abbasi, Adolfas K Gaigalas, Robert F Vogt, Gerald E Marti.   

Abstract

BACKGROUND: Numerous methods for quantitative fluorescence calibration (QFC) have been developed to quantify receptor expression on lymphocytes as potential disease biomarkers. CD20 expression in B-cell chronic lymphocytic leukemia (B-CLL) is one of the best examples of such a biomarker, but results from the use of different QFC methods vary considerably.
METHODS: We measured CD20 expression on normal and B-CLL B-cells, using FITC and PE conjugates from the same monoclonal antibody (Mab). As a biological control and calibrator, we also measured CD4 expression on T-cells with FITC and PE Mab. Calibration curves were constructed using the CLSI (formerly NCCLS) consensus guidelines for QFC. Calibration with QuantiBRITE PE-labeled microspheres and the use of unimolar PE conjugates provided direct measurement of antibody bound per cell (ABC) for CD4 and CD20. Calibration for FITC conjugates was based on molecules of equivalent soluble fluorochrome (MESF), as determined by NIST RM 8640 microsphere standards. These MESF values were then converted to ABC, using the CD4 T-cell as a biologic calibrator, to normalize FITC and PE results for CD20 expression.
RESULTS: On normal B cells, the mean ABC value for unimolar CD20-PE conjugate was 143,500 (CV +/- 19.1%). The mean ABC value for B-CLL B-cells stained with the same conjugate was 21,700 (CV +/- 42.0%). Using the CD4 T-cell as a biologic calibrator for FITC conjugate, the mean ABC value for CD20-FITC on normal B-cells was 199,300. CD20-FITC staining on B-CLL cells was generally too weak for accurate quantification. On normal T-cells, the mean ABC value for CD4 unimolar PE conjugate was (36,800 +/- 10.4)%, and it did not differ significantly in CLL samples.
CONCLUSION: The expression of CD20 on normal and B-CLL lymphocytes can be quantified in ABC units using unimolar CD20-PE conjugates. In addition, CD4 expression on T-cells can be used as a biological calibrator to quantify CD20-FITC ABC, with reasonable agreement between the two conjugates with different fluorochromes. Issues regarding the accuracy of MESF microsphere calibrators and effective F/P ratios for FITC conjugates will require additional laboratory studies. Copyright 2006 International Society for Analytical Cytology.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16967494     DOI: 10.1002/cyto.b.20140

Source DB:  PubMed          Journal:  Cytometry B Clin Cytom        ISSN: 1552-4949            Impact factor:   3.058


  11 in total

1.  Quantification and cell-to-cell variation of vascular endothelial growth factor receptors.

Authors:  P I Imoukhuede; Aleksander S Popel
Journal:  Exp Cell Res       Date:  2010-12-23       Impact factor: 3.905

2.  Antigen sensitivity of CD22-specific chimeric TCR is modulated by target epitope distance from the cell membrane.

Authors:  Scott E James; Philip D Greenberg; Michael C Jensen; Yukang Lin; Jinjuan Wang; Brian G Till; Andrew A Raubitschek; Stephen J Forman; Oliver W Press
Journal:  J Immunol       Date:  2008-05-15       Impact factor: 5.422

3.  Automated analysis of multidimensional flow cytometry data improves diagnostic accuracy between mantle cell lymphoma and small lymphocytic lymphoma.

Authors:  Habil Zare; Ali Bashashati; Robert Kridel; Nima Aghaeepour; Gholamreza Haffari; Joseph M Connors; Randy D Gascoyne; Arvind Gupta; Ryan R Brinkman; Andrew P Weng
Journal:  Am J Clin Pathol       Date:  2012-01       Impact factor: 2.493

4.  HuMab-7D8, a monoclonal antibody directed against the membrane-proximal small loop epitope of CD20 can effectively eliminate CD20 low expressing tumor cells that resist rituximab-mediated lysis.

Authors:  Tom van Meerten; Henk Rozemuller; Samantha Hol; Petra Moerer; Mieke Zwart; Anton Hagenbeek; Wendy J M Mackus; Paul W H I Parren; Jan G J van de Winkel; Saskia B Ebeling; Anton C Martens
Journal:  Haematologica       Date:  2010-09-17       Impact factor: 9.941

5.  Immunotherapy for B-Cell Neoplasms using T Cells expressing Chimeric Antigen Receptors: From antigen choice to clinical implementation.

Authors:  Mohamed-Rachid Boulassel; Ahmed Galal
Journal:  Sultan Qaboos Univ Med J       Date:  2012-07-15

6.  Quantification of expression of antigens targeted by antibody-based therapy in chronic lymphocytic leukemia.

Authors:  Prashant R Tembhare; Gerald Marti; Adrian Wiestner; Heba Degheidy; Mohammed Farooqui; Robert J Kreitman; Gregory A Jasper; Constance M Yuan; David Liewehr; David Venzon; Maryalice Stetler-Stevenson
Journal:  Am J Clin Pathol       Date:  2013-12       Impact factor: 2.493

7.  Development of Multicolor Flow Cytometry Calibration Standards: Assignment of Equivalent Reference Fluorophores (ERF) Unit.

Authors:  Lili Wang; Adolfas K Gaigalas
Journal:  J Res Natl Inst Stand Technol       Date:  2011-06-01

8.  Quantitative fluorescent profiling of VEGFRs reveals tumor cell and endothelial cell heterogeneity in breast cancer xenografts.

Authors:  Princess I Imoukhuede; Aleksander S Popel
Journal:  Cancer Med       Date:  2014-01-22       Impact factor: 4.452

9.  Expression of VEGF receptors on endothelial cells in mouse skeletal muscle.

Authors:  Princess I Imoukhuede; Aleksander S Popel
Journal:  PLoS One       Date:  2012-09-12       Impact factor: 3.240

10.  Flow Cytometry as a Tool for Quality Control of Fluorescent Conjugates Used in Immunoassays.

Authors:  Marta de Almeida Santiago; Bruna de Paula Fonseca E Fonseca; Christiane de Fátima da Silva Marques; Edimilson Domingos da Silva; Alvaro Luiz Bertho; Ana Cristina Martins de Almeida Nogueira
Journal:  PLoS One       Date:  2016-12-09       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.