| Literature DB >> 16967224 |
Daniel Giansante Abud1, Charbel Mounayer, Jean Pierre Saint-Maurice, Marco Túlio Salles Rezende, Emmanuel Houdart, Jacques Moret.
Abstract
We report 2 cases of hemorrhagic complications related to use of the Angio-Seal hemostatic closure device that were successfully managed with stent-grafts. Two patients with subarachnoid hemorrhage were referred to our departments for endovascular treatment of ruptured intracranial aneurysms. The treatment was performed through a femoral access; the sheaths were removed immediately after the procedures, and the punctures sites closed by Angio-Seals. Both patients presented clinical signs of hypovolemic shock after treatment. The diagnosis of active bleeding through the puncture site was made by emergency digital subtraction angiography. The lesions were managed with stent-grafts. The use of stent-grafts proved to be efficient in the management of these life-threatening hemorrhagic complications following the use of the Angio-Seal hemostatic closure device.Entities:
Mesh:
Year: 2007 PMID: 16967224 DOI: 10.1007/s00270-005-0321-8
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740