Quality assurance of I-125 seeds for prostate brachytherapy using an imaging plate.

Shunsuke Furutani1, Takuya Saze, Hitoshi Ikushima, Masataka Oita, Kyousuke Ozaki, Yoshiomi Kishida, Yoshihiro Takegawa, Hiromu Nishitani.   

Abstract

PURPOSE: OncoSeed is delivered in a sterile environment in the form of a cartridge, so it is impractical to resterilize and reload seeds after calibration. We investigated a new method using an imaging plate dosimetry system to characterize all seeds in the OncoSeed cartridge in a sterile environment. METHODS AND MATERIALS: Seeds within the cartridge were placed on an imaging plate, and the imaging plate irradiated. To remove scatter radiation, and improve spatial resolution of seed images, we used X-ray parallel cross grids. The irradiated imaging plate was scanned using a Bio-imaging Analyzer System, and radioactivity intensities of seed images were given in counts. Counts could be translated to profiles, and each seed within the cartridge was analyzed.
RESULTS: Results showed a good correlation between counts and total radioactivity of the seeds within the cartridge. Thus, using a least-squares line, it was possible to characterize a cartridge with unknown apparent activity. By analyzing the profiles, it was possible not only to detect a miscalibrated seed in the cartridge from its relative difference in counts, but also to identify its position in the cartridge. No significant changes in counts were seen between sterile and nonsterile environments.
CONCLUSION: Using an imaging plate dosimetry system, all seeds in a cartridge could be characterized in a sterile environment.

Mesh:

Substances:

Year:  2006        PMID: 16966003     DOI: 10.1016/j.ijrobp.2006.05.052

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  2 in total

1.  A method for confirming a third-party assay of I-125 seeds used for prostate implants.

Authors:  John S Muryn; D Allan Wilkinson
Journal:  J Appl Clin Med Phys       Date:  2016-11-24       Impact factor: 2.102

2.  Source strength assay of iodine-125 seeds sealed within sterile packaging.

Authors:  Yuki Otani; Takahiro Yamada; Shingo Kato; Naoto Shikama; Kazuto Funakoshi; Isao Kuroda; Hodaka Numasaki; Takayuki Nose; Takushi Dokiya; Masahiko Oguchi
Journal:  J Appl Clin Med Phys       Date:  2013-03-04       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.