Image-guided radiotherapy using megavoltage cone-beam computed tomography for treatment of paraspinous tumors in the presence of orthopedic hardware.

Eric K Hansen1, David A Larson, Michele Aubin, Josephine Chen, Martina Descovich, Amy M Gillis, Olivier Morin, Ping Xia, Jean Pouliot.   

Abstract

PURPOSE: This report describes a new image-guided radiotherapy (IGRT) technique using megavoltage cone-beam computed tomography (MV-CBCT) to treat paraspinous tumors in the presence of orthopedic hardware. METHODS AND MATERIALS: A patient with a resected paraspinous high-grade sarcoma was treated to 59.4 Gy with an IMRT plan. Daily MV-CBCT imaging was used to ensure accurate positioning. The displacement between MV-CBCT and planning CT images were determined daily and applied remotely to the treatment couch. The dose-volume histograms of the original and a hypothetical IMRT plan (shifted by the average daily setup errors) were compared to estimate the impact on dosimetry.
RESULTS: The mean setup corrections in the lateral, longitudinal, and vertical directions were 3.6 mm (95% CI, 2.6-4.6 mm), 4.1 mm (95% CI, 3.2-5.0 mm), and 1.0 mm (95% CI, 0.6-1.3 mm), respectively. Without corrected positioning, the dose to 0.1 cc of the spinal cord increased by 9.4 Gy, and the doses to 95% of clinical target volumes 1 and 2 were reduced by 4 Gy and 4.8 Gy, respectively.
CONCLUSIONS: Megavoltage-CBCT provides a new alternative image-guided radiotherapy approach for treatment of paraspinous tumors in the presence of orthopedic hardware by providing 3D anatomic information in the treatment position, with clear imaging of metallic objects and without compromising soft-tissue information.

Entities:  

Mesh:

Year:  2006        PMID: 16965986     DOI: 10.1016/j.ijrobp.2006.05.038

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  7 in total

1.  Monte Carlo investigations of megavoltage cone-beam CT using thick, segmented scintillating detectors for soft tissue visualization.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Amit Sawant; Hong Du
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

2.  Assessment and evaluation of MV image guidance system performance in radiotherapy.

Authors:  Nithya Kanakavelu; E James Jebaseelan Samuel
Journal:  Rep Pract Oncol Radiother       Date:  2015-02-07

3.  Dosimetry challenges for implementing emerging technologies.

Authors:  Fang-Fang Yin; Mark Oldham; Jing Cai; Qiuwen Wu
Journal:  J Phys Conf Ser       Date:  2010

4.  Prospective evaluation of spinal cord and cauda equina dose constraints using cone beam computed tomography (CBCT) image guidance for spine radiosurgery.

Authors:  Peter C Gerszten; Mubina Quader; Josef Novotny; John C Flickinger
Journal:  J Radiosurg SBRT       Date:  2011

5.  Image quality of an investigational imaging panel for use with the imaging beam line cone-beam CT.

Authors:  Chris Beltran
Journal:  J Appl Clin Med Phys       Date:  2012-01-05       Impact factor: 2.102

6.  An investigation into factors affecting electron density calibration for a megavoltage cone-beam CT system.

Authors:  Jessica Hughes; Lois C Holloway; Alexandra Quinn; Andrew Fielding
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

7.  Does the protocol-required uniform margin around the CTV adequately account for setup inaccuracies in whole breast irradiation?

Authors:  Jurui Luo; Zhihai Yin; Zhen Zhang; Xiaomao Guo; Xiaoli Yu; Juanqi Wang
Journal:  Radiat Oncol       Date:  2021-08-03       Impact factor: 3.481

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.