GLAaS: an absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications.

Giorgia Nicolini1, Antonella Fogliata, Eugenio Vanetti, Alessandro Clivio, Luca Cozzi.   

Abstract

A new calibration algorithm (GLAaS) to derive absolute dose maps from images acquired with the Varian PV-aS500 electronic portal imager (based on amorphous silicon detectors) has been developed incorporating the dependence of detector response on primary and transmitted radiation and on field size. Detector calibration and algorithm validation were performed at different depths (10.0, 3.8, 1.5, and 0.8 cm) in solid water to investigate various application possibilities. Calibration data were obtained against ion chamber measurements. Validation experiments were performed on intensity-modulated fields and comparison was carried out against calculated dose maps as well as against film measurements. Split fields were acquired independently and PV-aS500 images were summed offline with the new algorithm allowing complex fields to be verified in conditions most closely resembling clinical conditions. Excellent results were obtained for the 3.8, 1.5, and 0.8 depths on a set of 34 modulated fields including both split and nonsplit fields. Applying the gamma index analysis (with distance to agreement and dose thresholds set to 3 mm and 4%, respectively), only 2.3% of the field area showed gamma > 1 at 1.5 cm depth (8.1%, 3.1%, 2.7% at 10.0, 3.8, and 0.8 and 2.5% with films at 10 cm depth). Tests were also performed to verify GLAaS at gantry angles different from 0 degrees. No statistical differences were obtained for the comparison between split and nonsplit fields and between different gantry angles. Highly significant statistical differences were obtained when comparing independent samples of 240 fields verified either with GLAaS or with film. Fields verified with GLAaS presented a mean area with gamma > 1 of 2.1 +/-1.3% while for film this value was 3.9 +/- 3.4% (p<0.001). Absolute dosimetry proved to be reliable with the PV-aS500 detector with the GLAaS algorithm. The minimal settings at depths of 1.5 or 3.8 cm would allow the use of the detector at any gantry angle without the need for any special fixation tool.

Entities:  

Mesh:

Year:  2006        PMID: 16964860     DOI: 10.1118/1.2218314

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  Practical guidelines for routine intensity-modulated radiotherapy verification: pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry.

Authors:  A J Vinall; A J Williams; V E Currie; A Van Esch; D Huyskens
Journal:  Br J Radiol       Date:  2010-11       Impact factor: 3.039

2.  In vivo dosimetry in UK external beam radiotherapy: current and future usage.

Authors:  Niall D MacDougall; Michael Graveling; Vibeke N Hansen; Kevin Brownsword; Andrew Morgan
Journal:  Br J Radiol       Date:  2017-02-16       Impact factor: 3.039

3.  An empirical calibration method for an a-Si portal imaging device: applications in pretreatment verification of IMRT.

Authors:  L Conte; C Mordacchini; L Pozzi; C Vite
Journal:  Radiol Med       Date:  2012-03-19       Impact factor: 3.469

4.  Quality assurance of RapidArc in clinical practice using portal dosimetry.

Authors:  A Fogliata; A Clivio; P Fenoglietto; J Hrbacek; S Kloeck; P Lattuada; P Mancosu; G Nicolini; E Parietti; G Urso; E Vanetti; L Cozzi
Journal:  Br J Radiol       Date:  2011-06       Impact factor: 3.039

5.  Neo-adjuvant chemo-radiation of rectal cancer with volumetric modulated arc therapy: summary of technical and dosimetric features and early clinical experience.

Authors:  Antonella Richetti; Antonella Fogliata; Alessandro Clivio; Giorgia Nicolini; Gianfranco Pesce; Emanuela Salati; Eugenio Vanetti; Luca Cozzi
Journal:  Radiat Oncol       Date:  2010-02-19       Impact factor: 3.481

6.  Whole abdomen radiation therapy in ovarian cancers: a comparison between fixed beam and volumetric arc based intensity modulation.

Authors:  Umesh Mahantshetty; Swamidas Jamema; Reena Engineer; Deepak Deshpande; Rajiv Sarin; Antonella Fogliata; Giorgia Nicolini; Alessandro Clivio; Eugenio Vanetti; Shyamkishore Shrivastava; Luca Cozzi
Journal:  Radiat Oncol       Date:  2010-11-15       Impact factor: 3.481

7.  Large volume unresectable locally advanced non-small cell lung cancer: acute toxicity and initial outcome results with rapid arc.

Authors:  Marta Scorsetti; Pierina Navarria; Pietro Mancosu; Filippo Alongi; Simona Castiglioni; Raffaele Cavina; Luca Cozzi; Antonella Fogliata; Sara Pentimalli; Angelo Tozzi; Armando Santoro
Journal:  Radiat Oncol       Date:  2010-10-15       Impact factor: 3.481

8.  Early clinical experience with volumetric modulated arc therapy in head and neck cancer patients.

Authors:  Marta Scorsetti; Antonella Fogliata; Simona Castiglioni; Caterina Bressi; Mario Bignardi; Piera Navarria; Pietro Mancosu; Luca Cozzi; Sara Pentimalli; Filippo Alongi; Armando Santoro
Journal:  Radiat Oncol       Date:  2010-10-15       Impact factor: 3.481

9.  Early clinical experience of radiotherapy of prostate cancer with volumetric modulated arc therapy.

Authors:  Gianfranco A Pesce; Alessandro Clivio; Luca Cozzi; Giorgia Nicolini; Antonella Richetti; Emanuela Salati; Mariacarla Valli; Eugenio Vanetti; Antonella Fogliata
Journal:  Radiat Oncol       Date:  2010-06-16       Impact factor: 3.481

10.  Simultaneous integrated boost radiotherapy for bilateral breast: a treatment planning and dosimetric comparison for volumetric modulated arc and fixed field intensity modulated therapy.

Authors:  Giorgia Nicolini; Alessandro Clivio; Antonella Fogliata; Eugenio Vanetti; Luca Cozzi
Journal:  Radiat Oncol       Date:  2009-07-24       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.