Heparin plasma sampling as an alternative to EDTA for BNP determination on the Access-Beckman Coulter--effect of storage at -20 degrees C.

Anne Marie Dupuy1, Nathalie Terrier, Marjorie Dubois, Anne Marie Boularan, Jean Paul Cristol.   

Abstract

To determine BNP, EDTA plasma is the only suitable specimen recommended by the manufacturer. Since many laboratories, especially in Europe, use heparin plasma rather than EDTA plasma for many or most of their clinical assays and in particular for determination of cardiac markers (cTnI, myoglobin), it appeared critical to evaluate the use of heparin plasma samples, in comparison to EDTA plasma, for BNP determination on a automated immunochemiluminescent analyzer. The aim of this study was first, to evaluate the use of heparin plasma samples for Biosite BNP testing on the Beckman Coulter Access Immunoassay System (n=88) and second, to evaluate the effect of storage at -20 degrees C, without protease inhibitors, on the Biosite BNP assay. We obtained acceptable imprecision results with CVs ranging from 1.7 to 11.7% regardless of the anticoagulant used. The linearity of EDTA samples was good and comparable to the results observed with heparin plasma. The concentration of BNP was categorized according to the classification of the New York Heart Association (NYHA). With EDTA fresh samples as reference anticoagulant, 90% vs. 89% of subjects were classified as "concordant" with heparin fresh vs heparin frozen plasma samples, respectively. After storage at -20 degrees C, only 86% of the values of EDTA frozen were concordant with values of EDTA fresh. No subject varied by two NYHA classes. Heparin plasma is an attractive alternative to the established EDTA samples which can be used for BNP determination. This flexibility allows the simultaneous determination of CK, CK-MB, cTnI and BNP on a single heparin specimen, which facilitates blood collection for clinicians and nursing staff in an emergency unit. In addition, our results suggest that BNP could be stored at -20 degrees C for at least one month in order to perform retrospective studies.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16955639

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  2 in total

Review 1.  Pre-analytic variability in cardiovascular biomarker testing.

Authors:  Roberto Cemin; Massimo Daves
Journal:  J Thorac Dis       Date:  2015-10       Impact factor: 2.895

2.  Influence of sample matrix and storage on BNP measurement on the Bayer Advia Centaur.

Authors:  Giuseppe Lippi; Antonio Fortunato; Gian Luca Salvagno; Martina Montagnana; Giuliano Soffiati; Gian Cesare Guidi
Journal:  J Clin Lab Anal       Date:  2007       Impact factor: 2.352

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.