Intraoperative radiation therapy using a mobile electron linear accelerator: field matching for large-field electron irradiation.

A S Beddar1, T M Briere, M Ouzidane.   

Abstract

Intraoperative radiation therapy (IORT) consists of delivering a large, single-fraction dose of radiation to a surgically exposed tumour or tumour bed at the time of surgery. With the availability of a mobile linear accelerator in the OR, IORT procedures have become more feasible for medical centres and more accessible to cancer patients. Often the area requiring irradiation is larger than what the treatment applicators will allow, and therefore, two or more adjoining fields are used. Unfortunately, the divergence and scattering of the electron beams may cause significant dose variations in the region of the field junction. Furthermore, because IORT treatments are delivered in a large single fraction, the effects of underdosing or overdosing could be more critical when compared to fractionated external beam therapy. Proper matching of the fields is therefore an important technical aspect of treatment delivery. We have studied the matching region using the largest flat applicator available for three different possibilities: abutting the fields, leaving a small gap or creating an overlap. Measurements were done using film dosimetry for the available energies of 4, 6, 9 and 12 MeV. Our results show the presence of clinically significant cold spots for the low-energy beams when the fields are either gapped or abutted, suggesting that the fields should be overlapped. No fields should be gapped. The results suggest that an optimal dose distribution may be obtained by overlapping the fields at 4 and 6 MeV and simply abutting the fields at 9 and 12 MeV. However, due to uncertainties in the placement of lead shields during treatment delivery, one may wish to consider overlapping the higher energy fields as well.

Entities:  

Mesh:

Year:  2006        PMID: 16953035     DOI: 10.1088/0031-9155/51/18/N01

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Intraoperative implantation of a mesh of directional palladium sources (CivaSheet): Dosimetry verification, clinical commissioning, dose specification, and preliminary experience.

Authors:  Gil'ad N Cohen; Karen Episcopia; Seng-Boh Lim; Thomas J LoSasso; Mark J Rivard; Amandeep S Taggar; Neil K Taunk; Abraham J Wu; Antonio L Damato
Journal:  Brachytherapy       Date:  2017-08-18       Impact factor: 2.362

2.  Commissioning, clinical implementation, and performance of the Mobetron 2000 for intraoperative radiation therapy.

Authors:  Landon S Wootton; Juergen Meyer; Edward Kim; Mark Phillips
Journal:  J Appl Clin Med Phys       Date:  2017-01       Impact factor: 2.102

Review 3.  Present state and issues in IORT Physics.

Authors:  Frank W Hensley
Journal:  Radiat Oncol       Date:  2017-01-27       Impact factor: 3.481

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.