| Literature DB >> 16949889 |
Haihong Xu1, Wen Zhang, Dan Wang, Wei Zhu, Litong Jin.
Abstract
The fabrication and application of a novel electrochemical detection (ED) system with a poly(bromophenol blue) (PBPB) film chemically modified electrode (CME) for high performance liquid chromatography (HPLC) were described. The electrochemical behaviors of 5-hydroxytryptamine (5-HT) and 5-hydroxyindoleacetic acid (5-HIAA) at this CME were investigated by cyclic voltammetry (CV) and differential pulse voltammetry (DPV). It was found that the PBPB CME efficiently exhibited electrocatalytic effect on the current responses of 5-HT and 5-HIAA with relatively high sensitivity, stability and long life of activity. In HPLC-ED, the two analytes had good and stable current responses at the CME and their linear ranges were over four orders of magnitude (R> or =0.9992) with the detection limits being 0.25 nmol L(-1) for 5-HT and 0.50 nmol L(-1) for 5-HIAA. The application of this method for the determination of 5-HT and 5-HIAA in urine samples from patients with acute appendicitis (AA) was satisfactory.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16949889 DOI: 10.1016/j.jchromb.2006.08.021
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205