Experimental evaluation of ophthalmic devices and solutions using rabbit models.

Liliana Werner1, Jesse Chew, Nick Mamalis.   

Abstract

OBJECTIVE: To analyze and compare the geometry of the anterior segment of rabbit and human eyes, with relevance for the evaluation of intraocular lenses, and to review rabbit models used in our laboratory for the evaluation of different ophthalmic devices and solutions. PROCEDURES: Fifteen rabbit and 15 human eyes (10 phakic and 5 pseudophakic/group) obtained postmortem were used. Anterior-posterior length, equatorial diameter, and white-to-white (corneal diameter) were measured with calipers. The eyes were then analyzed with a very high-frequency ultrasound (Artemis, Ultralink) for measurements of the anterior chamber depth, and anterior chamber and ciliary sulcus diameters. The capsular bag diameter was measured with calipers from a posterior view, and the diameter and thickness of the crystalline lenses were measured after their excision from the phakic eyes.
RESULTS: Although the size of the rabbit eye is overall smaller than the size of the human eye, the dimensions of the anterior segment of rabbit eyes are generally larger. The differences between rabbit and human eyes were statistically significant (Wilcoxon rank sum test) in terms of anterior-posterior length, equatorial diameter, white-to-white measurements (P < 0.0001), anterior chamber diameter (P = 0.0004), ciliary sulcus diameter (P = 0.0012), and crystalline lens diameter and thickness (P = 0.0003).
CONCLUSIONS: Experimental evaluation of design features of new phakic intraocular lenses in rabbit eyes may be inconclusive without adaptation of their size/design, contrary to the evaluation of new pseudophakic lenses by implantation in the capsular bag. The rabbit is a very valuable model for the experimental evaluation of different ophthalmic devices and solutions.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16939455     DOI: 10.1111/j.1463-5224.2006.00495.x

Source DB:  PubMed          Journal:  Vet Ophthalmol        ISSN: 1463-5216            Impact factor:   1.644


  7 in total

1.  Thermal safety of ultrasound-enhanced ocular drug delivery: A modeling study.

Authors:  Marjan Nabili; Craig Geist; Vesna Zderic
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

2.  High incidence of spontaneous cataracts in aging laboratory rabbits of an inbred strain.

Authors:  Xuwen Peng; Sara Roshwalb; Timothy K Cooper; Heather Zimmerman; Neil D Christensen
Journal:  Vet Ophthalmol       Date:  2014-08-14       Impact factor: 1.644

3.  The precision of ophthalmic biometry using calipers.

Authors:  Ashik Mohamed; Derek Nankivil; Veerendranath Pesala; Mukesh Taneja
Journal:  Can J Ophthalmol       Date:  2013-12       Impact factor: 1.882

4.  Accelerated 20-year sunlight exposure simulation of a photochromic foldable intraocular lens in a rabbit model.

Authors:  Liliana Werner; Salwa Abdel-Aziz; Carolee Cutler Peck; Bryan Monson; Ladan Espandar; Brian Zaugg; Jack Stringham; Chris Wilcox; Nick Mamalis
Journal:  J Cataract Refract Surg       Date:  2011-02       Impact factor: 3.351

5.  Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.

Authors:  Dan Z Reinstein; Timothy J Archer; Marine Gobbe; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-11-16       Impact factor: 3.573

6.  Comparison of two prototypes of a magnetically adjustable glaucoma implant in rabbits.

Authors:  Birthe Schwerk; Lisa Harder; Claudia Windhövel; Marion Hewicker-Trautwein; Anna Wagner; Jan-Peter Bach; Lena Carolin Voigt; Ulf Hinze; Boris Chichkov; Heinz Haferkamp; Holger Lubatschowski; Stephan Nikolic; Ingo Nolte
Journal:  PLoS One       Date:  2019-04-11       Impact factor: 3.240

7.  A dose-escalation ex vivo study on the effects of intracameral benzalkonium chloride in rabbits.

Authors:  Sergi Segarra; Marta Leiva; Daniel Costa; Natàlia Coyo; Maria Sabés-Alsina; José Ríos; Teresa Peña
Journal:  BMC Vet Res       Date:  2018-02-02       Impact factor: 2.741

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.