Literature DB >> 16936463

Alteration of the pressure setting of a Codman-Hakim programmable valve by a television.

Satoshi Utsuki1, Satoru Shimizu, Hidehiro Oka, Sachio Suzuki, Kiyotaka Fujii.   

Abstract

A 7-year-old girl presented in semicomatose condition. She had received a ventriculoperitoneal shunt using a Codman-Hakim programmable valve. Head computed tomography demonstrated hydrocephalus and head radiography showed that the pressure setting of the shunt valve had changed to 60 mmH(2)O from 40 mmH(2)O. The pressure setting was returned to 40 mmH(2)O, and she was discharged because her clinical symptoms and hydrocephalus improved. One month later, she lost consciousness again and was transported to our hospital. Hydrocephalus and shunt valve pressure of 50 mmH(2)O were noted, and the pressure setting was returned to 40 mmH(2)O again. She was discharged without complications. We suspected that the valve pressure was caused by close contact with a television, because the patient tended to touch a television with her head during play. The valve pressure did not change after the television was placed on a high stand out of reach. We should recognize that there are many sources of weak magnetic fields that may influence a programmable valve in everyday life.

Entities:  

Mesh:

Year:  2006        PMID: 16936463     DOI: 10.2176/nmc.46.405

Source DB:  PubMed          Journal:  Neurol Med Chir (Tokyo)        ISSN: 0470-8105            Impact factor:   1.742


  3 in total

1.  Interactions between programmable shunt valves and the iPad 3 with Smart Cover.

Authors:  Yizheng He; Rory K J Murphy; Jarod L Roland; David D Limbrick
Journal:  Childs Nerv Syst       Date:  2013-02-20       Impact factor: 1.475

2.  Management of neonatal hydrocephalus: feasibility of use and safety of two programmable (Sophy and Polaris) valves.

Authors:  Juan F Martínez-Lage; María-José Almagro; Isabel Sanchez Del Rincón; Miguel A Pérez-Espejo; Claudio Piqueras; Raúl Alfaro; Javier Ros de San Pedro
Journal:  Childs Nerv Syst       Date:  2007-10-09       Impact factor: 1.475

3.  Maladjustment of programmable ventricular shunt valves by inadvertent exposure to a common hospital device.

Authors:  R Fujimura; R Lober; K Kamian; L Kleiner
Journal:  Surg Neurol Int       Date:  2018-03-01
  3 in total