Literature DB >> 16935831

Development and validation of RP-HPLC method for the analysis of metformin.

M Saeed Arayne1, Najma Sultana, M Hashim Zuberi.   

Abstract

The reversed-phase high-performance liquid chromatographic (RP-HPLC) method has been developed to quantify metformin hydrochloride (MfCl) in raw material and pharmaceutical formulations using C(18) analytical reverse-phase column. Diazepam was used as an internal standard. Mobile phase consisted of methanol-water (30:70 v/v), pumped at a flow rate of 0.5 ml/min at ambient temperature and the retention time was about 4.4 min with symmetrical peaks. (MfCl) was detected by ultraviolet absorbance at 233 nm with no interference of commonly used excipients. The method was linear over the concentration range 0.312-5 mug/mL (R2 = 0.9995). The limit of detection of metformin was 0.1 mug/mL and the limit of quantitation was 0.3 mug/mL. The results obtained showed a good agreement with the declared contents in case of pharmaceutical formulations. The proposed method is rapid, accurate, economical and selective and it may be used for the quantitative analysis of metformin in Neodipar tablets because of its sensitivity and reproducibility.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16935831

Source DB:  PubMed          Journal:  Pak J Pharm Sci        ISSN: 1011-601X            Impact factor:   0.684


  6 in total

1.  Transdermal Delivery of Metformin Using Dissolving Microneedles and Iontophoresis Patches for Browning Subcutaneous Adipose Tissue.

Authors:  Mehrnaz Abbasi; Zhaoyang Fan; John A Dawson; Shu Wang
Journal:  Pharmaceutics       Date:  2022-04-17       Impact factor: 6.525

2.  Spectrophotometric Quantitation of Metformin in Bulk Drug and Pharmaceutical Formulations using Multivariate Technique.

Authors:  M S Arayne; Najma Sultana; M H Zuberi; F A Siddiqui
Journal:  Indian J Pharm Sci       Date:  2009-05       Impact factor: 0.975

3.  Development and validation of reversed-phase high-performance liquid chromatography method for estimation of rizatriptan benzoate in oral strip formulations.

Authors:  S T Bhagawati; M Sreenivasa Reddy; Kiran Avadani; B S Muddukrishna; Swapnil J Dengale; Krishnamurthy Bhat
Journal:  J Basic Clin Pharm       Date:  2014-12

4.  HPLC method for estimation of metformin hydrochloride in formulated microspheres and tablet dosage form.

Authors:  Mousumi Kar; P K Choudhury
Journal:  Indian J Pharm Sci       Date:  2009-05       Impact factor: 0.975

5.  Development and Validation of a Simple and Sensitive LC-MS/MS Method for Quantification of Metformin in Dried Blood Spot Its Application as an Indicator for Medication Adherence.

Authors:  Bushra T ALquadeib; Nouf M Aloudah; Alanood S Almurshedi; Iman M ALfagih; Basmah N ALdosari; Adim S ALmeleky; Nour M Almubyedh
Journal:  Int J Gen Med       Date:  2021-07-08

6.  Hydrogel-forming microneedles enhance transdermal delivery of metformin hydrochloride.

Authors:  Eman M Migdadi; Aaron J Courtenay; Ismaiel A Tekko; Maelíosa T C McCrudden; Mary-Carmel Kearney; Emma McAlister; Helen O McCarthy; Ryan F Donnelly
Journal:  J Control Release       Date:  2018-07-07       Impact factor: 9.776

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.