Literature DB >> 16935291

Numerical and experimental studies on pulsatile flow in aneurysms arising laterally from a curved parent vessel at various angles.

Tong-Miin Liou1, Yi-Chen Li, Wei-Cheng Juan.   

Abstract

Both numerical and experimental studies have been performed to characterize the fluid flow inside the lateral aneurysms arising from the curved parent vessels at various angles gamma. The implicit solver was based on the time-dependent Navier-Stokes equations of incompressible laminar flow. Solutions were generated by a cell-center finite-volume method that used second order upwind and second order center flux difference splitting for the convection and diffusion term, respectively. The second order Crank-Nicolson method was used in the time integration term while the SIMPLEC algorithm was adopted to handle the pressure-velocity coupling. Complementarily, the particle tracking velocimetry (PTV) was used to measure the velocity fields. The conditions selected were to simulate an internal carotid artery with a diameter of 5 mm by similarity rules. The values of gamma explored were 0 degrees, 45 degrees, 90 degrees, and 135 degrees. Pulsatile flow with Wormersley number 3.9 and Reynolds numbers varying from 350 to 850 was considered. The computed results are firstly verified by the PTV measured ones. Discussion of the results is in terms of pulsatile main and secondary velocity vector fields, inflow rates into the aneurysm, and the distributions of wall shear stress and static pressure. It is found that among the angles examined gamma=45( composite function) is the riskiest angle from a fluid dynamics point of view and the aneurysmal dome is at risk.

Entities:  

Mesh:

Year:  2006        PMID: 16935291     DOI: 10.1016/j.jbiomech.2006.05.024

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  3 in total

1.  The 'Sphere': A Dedicated Bifurcation Aneurysm Flow-Diverter Device.

Authors:  Thomas Peach; J Frederick Cornhill; Anh Nguyen; Howard Riina; Yiannis Ventikos
Journal:  Cardiovasc Eng Technol       Date:  2014-08-26       Impact factor: 2.495

2.  Early experience in the treatment of intra-cranial aneurysms by endovascular flow diversion: a multicentre prospective study.

Authors:  James V Byrne; Radu Beltechi; Julia A Yarnold; Jacqueline Birks; Mudassar Kamran
Journal:  PLoS One       Date:  2010-09-02       Impact factor: 3.240

3.  Towards Predicting Patient-Specific Flow-Diverter Treatment Outcomes for Bifurcation Aneurysms: From Implantation Rehearsal to Virtual Angiograms.

Authors:  T W Peach; K Spranger; Y Ventikos