PURPOSE: To describe and to evaluate a novel application of magnetic resonance imaging (MRI) in determining the Ahmed glaucoma valve (AGV) function after implantation.
METHODS: Retrospective interventional case series of all surgical glaucoma patients in one practice were identified and orbital MRI series were performed. Orbital MRI findings were evaluated in all children who had undergone Ahmed glaucoma valve insertion.
RESULTS: A total of six patients were identified, ranging in age from 1 to 14 years. Mean follow-up time ranged from 1 to 3 years from the time of first tube insertion. Gadolinium enhancement MRIs showed an enhanced large cystic cavity over the plate in five patients, indicating partial to complete patency of the outflow of the tube. One MRI failed to show an enlarged enhanced bleb over the plate region, indicating potential obstruction of outflow. Surgical irrigation of the tube displayed obstruction. Ahmed tube exchange was performed. Enhanced cystic bleb showed compression of the extraocular muscles in four patients. These patients exhibited mild to moderate restrictive strabismus. No lenticular opacification was noted in the four phakic patients.
CONCLUSIONS: Early and late AGV failure and subsequent intraocular pressure rise are of great concern after pediatric glaucoma shunt surgery. Orbital MRI series should be considered in the cases of possible AGV drainage failure. In cases of reoperation, orbital MRI findings could be of great clinical value to evaluate for functional patency as well as for the EOM surgical intervention approach. To avoid unnecessary surgical exploration/explantation, orbital MRIs with gadolinium-dye enhancement should be considered in evaluating postshunt glaucoma surgical patients.

Entities:  

Mesh:

Year:  2006        PMID: 16935235     DOI: 10.1016/j.jaapos.2006.03.001

Source DB:  PubMed          Journal:  J AAPOS        ISSN: 1091-8531            Impact factor:   1.220


  5 in total

1.  Clinical findings following Ahmed Glaucoma Valve implantation in pediatric glaucoma.

Authors:  Amir Pirouzian; Joseph L Demer
Journal:  Clin Ophthalmol       Date:  2008-03

2.  A new manner of reporting pressure results after glaucoma surgery.

Authors:  Constantin-Dan Bordeianu; Cristina-Eugenia Ticu
Journal:  Clin Ophthalmol       Date:  2011-12-30

3.  Incidental Magnetic Resonance Imaging Findings of Postoperative Ocular Hypotony with Choroidal Effusion after Ahmed Glaucoma Valve Implantation: A Case Report.

Authors:  Hae Min Park; Hyo Chan Jeong; Won June Lee
Journal:  Korean J Ophthalmol       Date:  2022-05-06

4.  Assessment of Filtration Bleb and Endplate Positioning Using Magnetic Resonance Imaging in Eyes Implanted with Long-Tube Glaucoma Drainage Devices.

Authors:  Ichiya Sano; Masaki Tanito; Koji Uchida; Takashi Katsube; Hajime Kitagaki; Akihiro Ohira
Journal:  PLoS One       Date:  2015-12-07       Impact factor: 3.240

Review 5.  Magnetic Resonance Imaging Study on Blebs Morphology of Ahmed Valves.

Authors:  Joana Ferreira; Fernando Fernandes; Madalena Patricio; Ana Brás; Cristina Rios; Ingeborg Stalmans; Luis Abegão Pinto
Journal:  J Curr Glaucoma Pract       Date:  2015-01-15
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.