Literature DB >> 16930790

Locking plates for corrective osteotomy of malunited dorsally tilted distal radial fractures: a biomechanical study.

L P Müller1, D Klitscher, L Rudig, D Mehler, P M Rommens, K J Prommersberger.   

Abstract

The purpose of the study was to compare the biomechanical properties of five different palmar fixation plate designs in a distal radius osteotomy cadaver model. A 1cm metaphyseal osteotomy gap was made to simulate a corrective osteotomy and the osteotomy plated. Axial load was applied to the distal end of each construct by a material testing machine under control of a motion analysis video system. The specimens were arranged into five implant groups of eight specimens each. No test group developed deformity and movement of the fracture gap greater than 2mm with a load of 100N. Increasing the load to 250N revealed statistically significant differences in stiffness and failure load between the different plates. Axial failure strength and stiffness were greater for the radial correction plates than for the other implants. The former may provide enough stability for corrective osteotomy of dorsally angulated distal radial malunions, even when the osteotomy gap is only filled with cancellous bone graft instead of cortical bone graft.

Entities:  

Mesh:

Year:  2006        PMID: 16930790     DOI: 10.1016/j.jhsb.2006.06.002

Source DB:  PubMed          Journal:  J Hand Surg Br        ISSN: 0266-7681


  3 in total

1.  [Open wedge corrective osteotomy of malunited distal radius fractures through a palmar approach. A retrospective analysis].

Authors:  E Konul; H Krimmer
Journal:  Unfallchirurg       Date:  2012-07       Impact factor: 1.000

Review 2.  [Corrective surgery after fractures of the distal radius].

Authors:  K-J Prommersberger; J van Schoonhoven
Journal:  Unfallchirurg       Date:  2007-07       Impact factor: 1.000

3.  Failure of volar locking plate fixation of an extraarticular distal radius fracture: A case report.

Authors:  Jue Cao; Kagan Ozer
Journal:  Patient Saf Surg       Date:  2010-11-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.