Implant degradation and poor healing after endovascular repair of abdominal aortic aneurysms: an analysis of explanted stent-grafts.

Annie Major1, Robert Guidoin, Gilles Soulez, Louis A Gaboury, Guy Cloutier, Marc Sapoval, Yvan Douville, Guy Dionne, Robert H Geelkerken, Paul Petrasek, Sophie Lerouge.   

Abstract

PURPOSE: To study explanted stent-grafts to achieve a better understanding of the mechanisms of failure after endovascular treatment of abdominal aortic aneurysms (AAA).
METHODS: Twelve stent-grafts were harvested at autopsy (n=3) or during surgical conversion (n=9). Device alterations were investigated by macroscopic examination, radiography, and surface analysis techniques. Healing around the implants was studied via histology and immunohistochemistry, with particular attention to the stent-graft/tissue interface.
RESULTS: Degradation was more important with Vanguard stent-grafts (off the market) than with AneuRx and Talent stent-grafts, but rupture of nitinol wires and poor surface finish in Talent stent-grafts raise concern about their corrosion resistance and long-term stability. Poor healing was observed around stent-grafts even after several years of implantation, with absence of vascular smooth muscle cells, fibroblasts, and collagen formation. In addition to the well-known foreign body reaction around the graft, numerous polymorphonuclear cells characteristic of the first step of healing were present in tissues around stent-grafts retrieved at surgical conversion. Factors explaining the lack of tissue organization around stent-grafts are discussed.
CONCLUSION: The long-term stability of implants remains a concern and requires more transparency from manufacturers regarding the surface properties of their devices. Lack of neointima formation impairs biological fixation of the implant to the vessel wall, leading to possible endoleaks and migration. New-generation stent-grafts promoting biological fixation should be developed to improve clinical outcomes of this minimally invasive treatment.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16928159     DOI: 10.1583/06-1812MR.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  3 in total

1.  Comparative study of the corrosion behavior of peripheral stents in an accelerated corrosion model: experimental in vitro study of 28 metallic vascular endoprostheses.

Authors:  Karolin J Paprottka; Philipp M Paprottka; Maximilian F Reiser; Tobias Waggershauser
Journal:  Diagn Interv Radiol       Date:  2015 Sep-Oct       Impact factor: 2.630

Review 2.  The Potential Role of Graphene in Developing the Next Generation of Endomaterials.

Authors:  Nikolaos Patelis; Demetrios Moris; Sean Matheiken; Chris Klonaris
Journal:  Biomed Res Int       Date:  2016-11-29       Impact factor: 3.411

3.  An in Vitro Twist Fatigue Test of Fabric Stent-Grafts Supported by Z-Stents vs. Ringed Stents.

Authors:  Jing Lin; Robert Guidoin; Jia Du; Lu Wang; Graeham Douglas; Danjie Zhu; Mark Nutley; Lygia Perron; Ze Zhang; Yvan Douville
Journal:  Materials (Basel)       Date:  2016-02-16       Impact factor: 3.623

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.