Sac behavior after aneurysm treatment with the Gore Excluder low-permeability aortic endoprosthesis: 12-month comparison to the original Excluder device.

Shan-e-ali Haider1, Samer F Najjar, Jae-Sung Cho, Robert Y Rhee, Mark K Eskandari, Jon S Matsumura, Michel S Makaroun, Mark D Morasch.   

Abstract

BACKGROUND: The original Gore Excluder endoprosthesis (OGE) used both during and briefly after clinical trials was associated with less sac regression and more sac growth than some other devices, even without apparent endoleaks, presumably because of transmural movement of serous fluid across the expanded polytetrafluoroethylene material. In July 2004, the device was modified to decrease graft permeability. This study evaluated the efficacy of the new Excluder Low-Permeability Device (ELPD) at 1 year and compared it with the OGE and the Cook Zenith device (ZEN).
METHODS: From Food and Drug Administration approval of the Excluder in November 2002 until June 2005, 283 patients underwent endovascular repair of abdominal aortic aneurysms with the Gore Excluder or the ZEN. Postoperative surveillance included computed tomographic scans at 1 and 12 months; 181 (64%) patients completed both scans. The 1-month computed tomographic scan served as a baseline, and the minor axis diameter, measured at the largest axial cut of the abdominal aortic aneurysm, was compared with the same measurement at 1 year. A sac size change of 5 mm or more was considered significant. Sixty patients treated with the OGE were compared with 72 patients treated with the ELPD. Forty-nine patients treated during the same time period with the ZEN, known for early sac shrinkage, were used as a reference. All measurements were performed by one observer from a digital workstation. Wilcoxon signed rank tests (pairwise) or Kruskal-Wallis tests (three groups) were used for intergroup comparison of continuous variables, whereas chi2 statistics or Fisher exact tests were used to compare categorical variables.
RESULTS: Patient age and sex and mean maximum aneurysm diameter at baseline were similar among groups (P = .59, .27, and .46, respectively). Graft migration, stent fractures, acute surgical conversion, late abdominal aortic aneurysm rupture, or aneurysm-related deaths were not observed. Type II endoleak rates were similar between ELPD and ZEN (23.6% and 20.4%; P = .68). Although a higher rate of endoleaks was seen with OGE (36.7%), this was not significant when compared with the other two devices (P = .11). At 1 year, patients treated with ELPD had a sac regression rate that was significantly higher than that for patients treated with OGE (63.9% vs 25%; P < 0.001) and was similar to that for patients treated with ZEN (65.3%). Significant sac expansion was not observed with ELPD.
CONCLUSIONS: At 1 year, similar to ZEN, significant aneurysm sac regression and minimal sac expansion were noted after endovascular repair of abdominal aortic aneurysms with ELPD. Low-porosity fabric used in the construction of endoprostheses seems to be an important factor in early aneurysm sac shrinkage. Long-term efficacy regarding the prevention of sac enlargement remains unclear, and further follow-up is suggested.

Entities:  

Mesh:

Year:  2006        PMID: 16926082     DOI: 10.1016/j.jvs.2006.06.018

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  8 in total

1.  Successful endovascular management of an aortic rupture following stent placement for severe atherosclerotic stenosis: A case report.

Authors:  Thomas L Chung; Dipankar Mukherjee
Journal:  Int J Angiol       Date:  2007

2.  Endarterectomy of the aneurysm sac in open abdominal aortic aneurysm repair reduces perigraft seroma and improves graft incorporation.

Authors:  Thomas Wolff; Michael T Koller; Thomas Eugster; Christophe Rouden; Regula Marti; Lorenz Gürke; Peter Stierli
Journal:  World J Surg       Date:  2011-04       Impact factor: 3.352

3.  Clinical significance of endoleaks characterized by computed tomography during aortography performed immediately after endovascular abdominal aortic aneurysm repair: prediction of persistent endoleak.

Authors:  Motoki Nakai; Hirotatsu Sato; Morio Sato; Yuko Tanba; Yasutaka Noda; Akira Ikoma; Hiroki Sanda; Kohei Nakata; Hiroki Minamiguchi; Nobuyuki Kawai; Tetsuo Sonomura; Kazushi Kishi; Yosiharu Nishimura; Yoshitaka Okamura
Journal:  Jpn J Radiol       Date:  2012-10-10       Impact factor: 2.374

4.  Endovascular treatment of perigraft seroma after open abdominal aortic aneurysm repair: report of a case.

Authors:  Kenji Sangawa; Atsushi Aoki
Journal:  Ann Vasc Dis       Date:  2013-05-30

5.  Low-permeability Gore Excluder device versus the original in abdominal aortic aneurysm size regression.

Authors:  Zvonimir Krajcer; Lawrence Rajan; Venkata Thota; Kathryn G Dougherty; Neil E Strickman; Ali Mortazavi; Benjamin Y C Cheong
Journal:  Tex Heart Inst J       Date:  2011

6.  A randomized, placebo-controlled trial of doxycycline after endoluminal aneurysm repair.

Authors:  Amy E Hackmann; Brian G Rubin; Luis A Sanchez; Patrick A Geraghty; Robert W Thompson; John A Curci
Journal:  J Vasc Surg       Date:  2008-07-15       Impact factor: 4.268

Review 7.  Endotension: twenty years of a controversial term.

Authors:  Álvaro Torres-Blanco; Manuel Miralles-Hernández
Journal:  CVIR Endovasc       Date:  2021-06-05

8.  Statins Reduce Abdominal Aortic Aneurysm Growth, Rupture, and Perioperative Mortality: A Systematic Review and Meta-Analysis.

Authors:  Konrad Salata; Muzammil Syed; Mohamad A Hussain; Charles de Mestral; Elisa Greco; Muhammad Mamdani; Jack V Tu; Thomas L Forbes; Deepak L Bhatt; Subodh Verma; Mohammed Al-Omran
Journal:  J Am Heart Assoc       Date:  2018-10-02       Impact factor: 5.501

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.