Multi-center clinical experience with a lumenless, catheter-delivered, bipolar, permanent pacemaker lead: implant safety and electrical performance.

Michael D Gammage1, Randy A Lieberman, Raymond Yee, Antonis S Manolis, Steven J Compton, César Khazen, Katie Schaaf, Kimberly A Oleson, George H Crossley.   

Abstract

PURPOSE: Reduced lead diameter and reliability can be designed into transvenous permanent pacing leads through use of redundant insulation and removal of the stylet lumen. The model 3830 lead (Medtronic Inc., Minneapolis, MN, USA) is a bipolar, fixed-screw, steroid-eluting, lumenless, 4.1-Fr pacing lead. Implantation can be performed in a variety of right heart sites using a deflectable catheter (Model 10600, Medtronic). Lead performance and safety were studied.
METHODS: Two prospective trials of 338 implanted subjects from 56 global sites were conducted. Electrical and safety data were obtained at implant, pre-discharge, and up to 18 months post-implant. Leads were implanted at traditional and alternate right heart sites.
RESULTS: The study enrolled 338 subjects (204 males, 70.6 +/- 11.6 years) followed-up for a mean of 10.2 months (range, 0-21.6). Mean P-wave amplitudes ranged from 3.2 mV at 3 months to 2.9 mV at 18 months, while mean atrial pulse width thresholds at 2.5 V ranged from 0.07 ms at 3 months to 0.09 ms at 18 months. Mean R-wave amplitudes ranged from 11.3 mV to 11.1 mV and mean ventricular pulse width thresholds at 2.5 V ranged from 0.10 ms to 0.14 ms. There were 22 ventricular and 12 atrial lead complications within 3 months post-implant. Survival from lead-related complications improved to a clinically acceptable rate in the cohort of patients when revised implant techniques were employed.
CONCLUSIONS: With the use of recommended implant techniques, the study results support the electrical efficacy and safety of a catheter-delivered, lumenless lead in traditional or alternate right atrium or right ventricle sites through 18 months post-implant.

Entities:  

Mesh:

Year:  2006        PMID: 16923002     DOI: 10.1111/j.1540-8159.2006.00452.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  11 in total

1.  In Vitro/Ex Vivo Investigation of Modified Utah Electrode Array to Selectively Sense and Pace the Sub-Surface Cardiac His Bundle.

Authors:  Ankur R Shah; Muhammad S Khan; Annie M Hirahara; Matthias Lange; Ravi Ranjan; Derek J Dosdall
Journal:  ACS Biomater Sci Eng       Date:  2020-05-07

2.  Implantation of lumenless pacing leads at the inter-atrial septum and right ventricular outflow tract with deflectable catheter-sheath.

Authors:  Rong Bai; Ruth Kam; Chi Keong Ching; Li Fern Hsu; Wee Siong Teo
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2008-12-24

3.  An ergonomic guide catheter slitting technique designed to avoid lead dislodgement.

Authors:  Ernest W Lau
Journal:  J Interv Card Electrophysiol       Date:  2008-06-06       Impact factor: 1.900

Review 4.  Should His Bundle Pacing Be Preferred over Cardiac Resynchronization Therapy Following Atrioventricular Junction Ablation?

Authors:  Zak Loring; Albert Y Sun
Journal:  Cardiol Clin       Date:  2019-02-10       Impact factor: 2.213

5.  A streamlined technique of trans-septal endocardial left ventricular lead placement.

Authors:  Ernest W Lau
Journal:  J Interv Card Electrophysiol       Date:  2009-04-22       Impact factor: 1.900

6.  The Safety and Feasibility of Same-Day Discharge After Implantation of MICRA Transcatheter Leadless Pacemaker System.

Authors:  Soroosh Kiani; George B Black; Birju Rao; Nancy Thakkar; Christopher Massad; Akshar V Patel; Marvin Louis Roy Lu; Faisal M Merchant; Michael H Hoskins; David B De Lurgio; Anshul M Patel; Anand D Shah; Angel R Leon; Stacy B Westerman; Michael S Lloyd; Mikhael F El-Chami
Journal:  J Atr Fibrillation       Date:  2019-06-30

7.  Effects of flecainide and quinidine on arrhythmogenic properties of Scn5a+/- murine hearts modelling the Brugada syndrome.

Authors:  Kate S Stokoe; Richard Balasubramaniam; Catharine A Goddard; William H Colledge; Andrew A Grace; Christopher L-H Huang
Journal:  J Physiol       Date:  2007-02-15       Impact factor: 5.182

8.  Right ventricular pacing and sensing function in high posterior septal and apical lead placement in cardiac resynchronization therapy.

Authors:  Hm Kristiansen; T Hovstad; G Vollan; S Faerestrand
Journal:  Indian Pacing Electrophysiol J       Date:  2012-01-31

9.  Case series of late lead dislodgement of Medtronic SelectSecure 3830 pacing leads in growing paediatric patients.

Authors:  Li Yen Ng; Sarah Gallagher; Kevin P Walsh
Journal:  Eur Heart J Case Rep       Date:  2021-01-12

10.  Effects of flecainide and quinidine on arrhythmogenic properties of Scn5a+/Delta murine hearts modelling long QT syndrome 3.

Authors:  Kate S Stokoe; Glyn Thomas; Catharine A Goddard; William H Colledge; Andrew A Grace; Christopher L-H Huang
Journal:  J Physiol       Date:  2006-10-05       Impact factor: 5.182

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.