Literature DB >> 16914305

Surface modified amorphous ribbon based magnetoimpedance biosensor.

Galina V Kurlyandskaya1, Vanessa Fal Miyar.   

Abstract

Magnetoimpedance (MI) changes due to surface modification of the sensitive element caused by human urine, were studied with the aim of creating a robust biosensor working on a principle of electrochemical magnetoimpedance spectroscopy. A biosensor prototype with an as-quenched amorphous ribbon sensitive element was designed and calibrated for a frequency range of 0.5-10 MHz at a current intensity of 60 mA. Measurements as a function of the exposure time were made both in a regime where chemical surface modification and MI measurements were separated as well as in a regime where they were done simultaneously. The MI variation was explained by the change of the surface magnetic anisotropy. It was shown that the magnetoimpedance effect can be successfully employed as a new option to probe the electric features of the Fe(5)Co(70)Si(15)B(10) amorphous ribbon magnetic electrode surface modified by human urine.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16914305     DOI: 10.1016/j.bios.2006.07.011

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  2 in total

1.  Low-cost Sensors Based on the GMI Effect in Recycled Transformer Cores.

Authors:  Pongsakorn Jantaratana; Chitnarong Sirisathitkul
Journal:  Sensors (Basel)       Date:  2008-03-10       Impact factor: 3.576

Review 2.  Magnetoimpedance Biosensors and Real-Time Healthcare Monitors: Progress, Opportunities, and Challenges.

Authors:  Valery Ortiz Jimenez; Kee Young Hwang; Dang Nguyen; Yasif Rahman; Claire Albrecht; Baylee Senator; Ongard Thiabgoh; Jagannath Devkota; Vinh Duc An Bui; Dao Son Lam; Tatiana Eggers; Manh-Huong Phan
Journal:  Biosensors (Basel)       Date:  2022-07-12
  2 in total