Literature DB >> 16912369

Downscatter correction and choice of collimator in 123I imaging.

A D Small1, J Prosser, D W Motherwell, G M McCurrach, A M Fletcher, W Martin.   

Abstract

(123)I imaging is increasingly used in nuclear medicine but downscatter from high (>300 keV) energy emissions degrades the image and introduces variation in sensitivity with depth when imaging with a low energy collimator. A dual windowing technique using a matched window immediately above the principle window to correct for the effects of downscatter is described. The technique is shown to correct for the variation in sensitivity with depth and to reduce outlying scatter. Quantitative imaging of (123)I using a low energy collimator should always employ downscatter correction.

Mesh:

Substances:

Year:  2006        PMID: 16912369     DOI: 10.1088/0031-9155/51/17/N03

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Quantification of I-123-meta-iodobenzylguanidine heart-to-mediastinum ratios: not so simple after all.

Authors:  Piotr J Slomka; Puja K Mehta; Guido Germano; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2014-07-09       Impact factor: 5.952

2.  Experimental determination of the weighting factor for the energy window subtraction-based downscatter correction for I-123 in brain SPECT studies.

Authors:  Robin de Nijs; Søren Holm; Gerda Thomsen; Morten Ziebell; Claus Svarer
Journal:  J Med Phys       Date:  2010-10

3.  Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections.

Authors:  Robin de Nijs; Vera Lagerburg; Thomas L Klausen; Søren Holm
Journal:  Nucl Med Commun       Date:  2014-05       Impact factor: 1.690

  3 in total