In-vivo validation of spatially correct three-dimensional reconstruction of human coronary arteries by integrating intravascular ultrasound and biplane angiography.

George D Giannoglou1, Yiannis S Chatzizisis, George Sianos, Dimitrios Tsikaderis, Antonis Matakos, Vassilios Koutkias, Panagiotis Diamantopoulos, Nicos Maglaveras, George E Parcharidis, George E Louridas.   

Abstract

OBJECTIVES: The in-vivo validation of geometrically correct three-dimensional reconstruction of human coronary arteries by integrating intravascular ultrasound and biplane coronary angiography has not been adequately investigated. The purpose of this study was to describe the reconstruction method and investigate its in-vivo feasibility and accuracy.
METHODS: In 17 coronary arteries (mean length, 85.7+/-17.1 mm) from nine patients, an intravascular ultrasound procedure along with a biplane coronary angiography was performed. From each angiographic projection, a single end-diastolic frame was selected in order to reconstruct the intravascular ultrasound catheter trajectory in space. In each end-diastolic intravascular ultrasound image, the lumen and media-adventitia contours were detected semi-automatically by an active contour algorithm. Each pair of contours was located on the catheter trajectory appropriately and interpolated with the adjacent pairs creating a three-dimensional volume of the arterial lumen and wall. The reconstructed lumen was back-projected onto both angiographic planes and the agreement between the back-projected and the angiographic luminal outlines was calculated.
RESULTS: The angiogram-derived catheter length showed very high correlation (y=0.97 x + 1.8, P<0.001) and agreement with the corresponding pullback-derived values. Accordingly, the semi-automated segmentation of intravascular ultrasound images was also in significant correlation (r> or =0.96, P<0.001) and agreement with the reference manual tracing. The back-projected luminal borders showed good overall association with the corresponding angiographic ones (r=0.78, P<0.001) as well as remarkable agreement.
CONCLUSIONS: Spatially correct three-dimensional reconstruction of human coronary arteries constitutes an imaging method with considerably high in-vivo feasibility and accuracy.

Entities:  

Mesh:

Year:  2006        PMID: 16905966     DOI: 10.1097/00019501-200609000-00007

Source DB:  PubMed          Journal:  Coron Artery Dis        ISSN: 0954-6928            Impact factor:   1.439


  3 in total

1.  Texture Analysis and Radial Basis Function Approximation for IVUS Image Segmentation.

Authors:  Maria Papadogiorgaki; Vasileios Mezaris; Yiannis S Chatzizisis; George D Giannoglou; Ioannis Kompatsiaris
Journal:  Open Biomed Eng J       Date:  2007-09-28

2.  A model of blood flow in the mesenteric arterial system.

Authors:  Thusitha D S Mabotuwana; Leo K Cheng; Andrew J Pullan
Journal:  Biomed Eng Online       Date:  2007-05-08       Impact factor: 2.819

Review 3.  The Evolution of Data Fusion Methodologies Developed to Reconstruct Coronary Artery Geometry From Intravascular Imaging and Coronary Angiography Data: A Comprehensive Review.

Authors:  Yakup Kilic; Hannah Safi; Retesh Bajaj; Patrick W Serruys; Pieter Kitslaar; Anantharaman Ramasamy; Vincenzo Tufaro; Yoshinobu Onuma; Anthony Mathur; Ryo Torii; Andreas Baumbach; Christos V Bourantas
Journal:  Front Cardiovasc Med       Date:  2020-03-31
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.