Daisuke Inui1, Jun Oto, Masaji Nishimura.
Abstract
BACKGROUND: In mechanically ventilated patients, we investigated how positioning the heat and moisture exchanger (HME) at different places on the ventilator circuit affected inspiratory gas humidification.Entities:
Mesh:
Year: 2006 PMID: 16895607 PMCID: PMC1562445 DOI: 10.1186/1471-2466-6-19
Source DB: PubMed Journal: BMC Pulm Med ISSN: 1471-2466 Impact factor: 3.317
Figure 1A heat and moisture exchanger (HME) was connected directly to the ETT (Site 1) and to the intersection of the Y piece and a 10 cm length of corrugated tube (Site 2).
Patient profile
| No. | Diagnosis | Gender | Age (y) | BW (kg) | BT(°C) | RT(°C) | Vt(mL) | MV (L) |
| 1 | cavernous angiomapo. | M | 32 | 60 | 36.8 | 26.2 | 600 | 8.1 |
| 2 | acute bronchitis | F | 26 | 29 | 36.0 | 27.0 | 230 | 2.8 |
| 3 | hypertrophic CM | M | 12 | 70 | 38.0 | 29.1 | 450 | 8.1 |
| 4 | gastric cancer po. | M | 74 | 75 | 37.4 | 29.7 | 530 | 9.1 |
| 5 | PE po. | M | 55 | 55 | 36.3 | 26.1 | 700 | 11.3 |
| 6 | AAA po. | M | 56 | 62 | 35.0 | 26.0 | 550 | 7.0 |
| 7 | IA po. | M | 72 | 72 | 37.2 | 26.0 | 520 | 7.8 |
| 8 | pneumonia | M | 68 | 55 | 36.6 | 27.0 | 260 | 8.7 |
| 9 | meningitis | M | 74 | 51 | 36.8 | 26.0 | 553 | 10.3 |
| 10 | AR po. | M | 42 | 61 | 36.5 | 26.2 | 380 | 9.0 |
| Mean | 51.1 | 59.0 | 36.7 | 26.9 | 477.3 | 8.2 |
po. postoperation, CM: cardiomyopathy, PE: pulmonary emobolism, AAA: abdominal aortic aneurysm, IA: intracranial aneurysm, AR: aortic regurgitation, M: male, F:female, BW: body weight, BT: body temperature, RT: room temperature, Vt: tidal volume, MV: minute volume
Figure 2Absolute humidity (AH) with Hygrobac S or Thermovent HEPA placed at Site 1 or Site 2. AH was significantly higher when Hygrobac S (closed circle)(*p < 0.05) was placed at Site 1 than at Site 2, and when Thermovent HEPA (open circle)(**p < 0.05) was placed at Site 1 than at Site 2. Patient N = 10. Data presented as mean ± SD.
Figure 3Temperature (TEMP) with Hygrobac S or Thermovent HEPA placed at Site 1 or Site 2. TEMP was significantly higher when Hygrobac S (closed circle)(*p < 0.05) was placed at Site 1 than at Site 2, and when Thermovent HEPA (open circle) was (**p < 0.05) was placed at Site 1 than at Site 2. Patient N = 10. Data presented as mean ± SD.